Selection
The chiller is intended only for externally closed applications, so the process loop must be closed and not open to atmosphere.
Determines whether the unit can be used with an open bath or only with a closed external circuit.
Source
“For externally closed applications.”
Installation
The unit requires a three-phase 400 V 3~ 50 Hz supply, so site power must be three-phase.
Electrical supply must be verified before ordering and installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and install the plug.
Affects electrical installation scope and requires a suitable plug on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and some combinations are not allowed.
Supply quality must be checked; out-of-tolerance combinations can invalidate operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit must be installed where ambient temperature stays between 5 °C and 40 °C.
Ambient conditions affect cooling capacity and safe operation.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises.
Selection must account for actual site ambient temperature, not just nominal capacity.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The pump connection is G1 1/4 male, so mating hoses or fittings must match this thread.
Ensures correct process plumbing connection.
Source
“Pump connection G1 1/4 male”
Installation
Space requirements must be respected; dimensions are 600x790x1614 mm and the unit weighs 230 kg.
Needed for floor loading, access, and clearance planning.
Source
“Please respect space requirements.”
Safety
The refrigerant is R-449A with GHS classification A1, H280, and GWP 1397.
Refrigerant safety and environmental data are needed for risk assessment and compliance.
Source
“R-449A (A1, H280)”
Cumplimiento
The equipment is EMC Class A, Group 1, which may require additional measures in residential or sensitive environments.
Affects installation location and EMC compatibility with nearby equipment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation
The unit has IP20 protection, so it must be protected from water and dust ingress.
Determines suitable environment and cleaning/installation precautions.
Source
“Degree of Protection IP20”
Maintenance
The expansion tank is atmospheric open and has a cover as standard equipment, which must be handled correctly during operation.
Open expansion tank design affects filling, venting, and maintenance.
Source
“cover expansion tank*”