Unichiller 130T-H Technical Data Sheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller 130T-H chiller model
  2. Spec pp. 1-2
    Performance, temperature, and physical specifications per DIN 12876
  3. p. 2 table, 6 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/19

Technical data according to DIN 12876 — Unichiller 130T-H (Order-No.: 3018.0018.01)

CharacteristicValue
Operating temperature range-10...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C13 kW
Cooling power at 0°C7 kW
Cooling power at -10°C4,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:C3
max. delivery90 l/min
max. delivery pressure5,65 bar
Delivery at 0,33 bar94 l/min
Delivery at 0,5 bar92 l/min
Delivery at 1,0 bar87 l/min
Delivery at 1,5 bar82 l/min
Delivery at 2,0 bar75,5 l/min
Delivery at 2,5 bar69 l/min
Delivery at 3,0 bar62 l/min
Delivery at 3,5 bar54,5 l/min
Delivery at 4,0 bar47 l/min
Delivery at 4,5 bar38 l/min
Delivery at 5,0 bar26 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **904x1582x1837 mm
sound pressure level +/- 4 dB(A)74 dB(A)

** см. сноску в оригинале (Overall dimensions WxDxH **)

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply, so the site must provide a 3~ 50 Hz feed.

Determines electrical infrastructure and whether a suitable three-phase supply is available at the installation point.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the plug.

Affects electrical installation planning and the need to procure a compatible plug or hard-wired connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts installation environment and rules out washdown or outdoor locations.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C for valid operation.

Defines the allowable room conditions and warns that cooling capacity may drop if ambient rises.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and may decrease at higher ambient temperatures.

Prevents oversizing assumptions and requires derating when the installation room is warmer than 20 °C.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but not all combinations are allowed.

Guides power quality assessment and avoids invalid supply combinations that could damage the unit.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Cumplimiento

The equipment emits Class A, Group 1 disturbances per EN55011, so EMC measures may be needed in residential-like areas.

Affects site EMC planning and suitability for sensitive or residential environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The refrigerant is R-449A with GHS classification H280, so refrigerant handling and ventilation precautions apply.

Informs safety planning for refrigerant storage, leak response, and room ventilation.

Source
“R-449A (A1, H280)”
Selection

The pump connection is G1 1/4 male, so external piping must match this thread.

Determines the hose couplings and adapters needed for the external circuit.

Source
“Pump connection G1 1/4 male”
Selection

The minimum filling capacity is 14 l and the expansion volume is 39 l, which sets the external loop volume limits.

Helps size the external circuit and expansion tank requirements.

Source
“min. filling capacity 14 l Volume of expansion 39 l”
Installation

The unit is air-cooled and requires respect of space requirements around it.

Ensures adequate airflow and service access, which affects cooling performance and maintenance.

Source
“Please respect space requirements.”
Operation

The unit is intended for externally closed applications, so the external circuit must be closed.

Prevents use with open baths or unsealed loops that could cause contamination or overflow.

Source
“For externally closed applications.”