Installation
The chiller requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the mains connection, cable and protection device needed at the installation site.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide and fit the plug.
Affects site preparation and the scope of electrical works before commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit must be installed in an environment between 5 °C and 40 °C ambient temperature.
Room conditions must be checked during selection and site planning.
Source
“min. ambient temperature 5 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and may drop if the ambient temperature rises.
The specified 13 kW at 15 °C is not guaranteed at higher ambient temperatures, so selection must account for the actual room temperature.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The enclosure is IP20, so it must be protected from water and dust ingress at the installation location.
Restricts where the chiller can be installed and what environmental protection is needed.
Source
“Degree of Protection IP20”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation requirements apply.
Affects safety planning, leak detection and personnel qualification for service.
Source
“R-449A (A1, H280)”
Cumplimiento
The CO2 equivalent of the refrigerant charge is 6.1 t, which triggers F-gas record-keeping and leak-check obligations in many jurisdictions.
Determines regulatory obligations for operation and maintenance of the refrigeration circuit.
Source
“CO2 equivalent 6,1 t”
Installation
The pump connection is G1 1/4 male, so the external circuit piping must match this thread.
Needed to order the correct hose couplings and plan the hydraulic connection.
Source
“Pump connection G1 1/4 male”
Operation
The minimum filling capacity is 14 l and the expansion volume is 39 l, which sets the minimum external loop volume and expansion allowance.
The external circuit must provide at least this volume and accommodate expansion to avoid operating problems.
Source
“min. filling capacity 14 l”
Installation
The unit is air-cooled and requires space around it, so the stated dimensions and space requirements must be respected.
Insufficient clearance reduces cooling performance and can cause overheating.
Source
“Please respect space requirements.”
Cumplimiento
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
The site power quality must be checked to avoid invalid supply conditions and equipment damage.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The sound pressure level is 74 dB(A), which must be considered for personnel exposure and room acoustics.
Affects workplace noise assessment and installation location.
Source
“sound pressure level +/- 4 dB(A) 74 dB(A)”
Selection
The temperature stability at -10 °C is 0.5 K, which defines the achievable process control accuracy at low temperature.
Used to verify whether the chiller meets the process tolerance requirements.
Source
“Temperature stability at -10°C 0,5 K”
Maintenance
A 4-year warranty is available only with registration, so the unit must be registered after purchase.
Impacts lifecycle cost and the need to complete registration at commissioning.
Source
“4-year warranty - registration required.”