The chiller is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.
Defines the application window and excludes processes outside this range.
Source
“Operating temperature range -20...100 °C”
| Parameter | Value |
|---|---|
| Net weight | 476 kg |
| Power supply (3 Phase) | 400V 3~ 50Hz |
| max. current (3 Phase) | 14,5 A |
| Fuse (3 phase) | 3x16A |
| Pressure equipment category | I |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | S590337 |
| Parameter | Value |
|---|---|
| Operating temperature range | -20...100 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Temperature stability at -10°C | 0,5 K |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Safety classification | III / FL |
| Heating power | 2 kW |
| Cooling power at ambient temperature +20°C | |
| at 100°C | 18 kW |
| at 20°C | 18 kW |
| at 0°C | 10 kW |
| at -10°C | 6 kW |
| at -20°C | 3,5 kW |
| Refrigeration machine | air-cooled, CFC- and HCFC-free |
| Refrigerant (ASHRAE, GHS) | R-449A (A1, H280) |
| Global Warming Potential (GWP) | 1397 |
| Refrigerant quantity | 1,7 kg |
| CO2 equivalent | 2,4 t |
| UN-number | UN 2857 |
| Circulation pump: | E1 |
| max. delivery | 54 l/min |
| max. delivery pressure | 3,5 bar |
| Delivery at 0,3 bar | 52 l/min |
| Delivery at 0,5 bar | 48 l/min |
| Delivery at 1,0 bar | 42 l/min |
| Delivery at 1,5 bar | 36 l/min |
| Delivery at 2,0 bar | 28 l/min |
| Delivery at 2,5 bar | 21 l/min |
| Delivery at 3,0 bar | 11 l/min |
| Pump connection | G1 1/4 male |
| min. filling capacity | 18 l |
| Volume of expansion | 48 l |
| Overall dimensions WxDxH ** | 938x1288x2003 mm |
** см. сноску в оригинале (на странице не приведена).
The chiller is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.
Defines the application window and excludes processes outside this range.
“Operating temperature range -20...100 °C”
Cooling capacity drops sharply at low temperatures (18 kW at 20 °C vs 3.5 kW at -20 °C), so size the unit for the coldest set point.
Prevents undersizing when the process requires low-temperature cooling.
“at -20°C 3,5 kW”
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical infrastructure must be prepared accordingly.
Determines the site power connection and protection device rating.
“Power supply (3 Phase) 400V 3~ 50Hz”
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the plug.
Affects electrical installation planning and bill of materials.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Voltage and frequency tolerances are interdependent: -5% voltage with +2% frequency is not allowed, so the site supply must be checked.
Avoids invalid supply conditions that could damage the compressor or void warranty.
“-5% voltage and +2% frequency -> not allowed!”
The unit is rated for ambient temperatures from 5 to 40 °C, and cooling capacity may drop above 20 °C.
Room conditions must be controlled to maintain rated performance.
“If the ambient temperature rises, the cooling capacity may drop.”
The refrigerant is R-449A with GHS hazard H280 (pressurized gas), so handling and ventilation precautions apply.
Informs safety measures and refrigerant handling requirements.
“R-449A (A1, H280)”
The equipment emits Class A, Group 1 disturbances per EN55011, so it is intended for industrial environments.
Restricts use in residential or light-industrial EMC environments.
“Classification (disturbance) to EN55011: Class A, Group 1”
The unit has IP20 protection, so it must be installed in a dry, indoor location protected from water and dust ingress.
Determines suitable installation environment and enclosure needs.
“Degree of Protection IP20”
The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread.
Ensures correct hydraulic connection and avoids adapters.
“Pump connection G1 1/4 male”
The minimum filling capacity is 18 l and expansion volume is 48 l, so the external loop volume must be considered.
Prevents insufficient fill or overflow in closed external circuits.
“min. filling capacity 18 l”
The unit is intended for externally closed applications, so open-loop or tank applications are not supported.
Confirms the hydraulic circuit type the chiller is designed for.
“For externally closed applications.”
The refrigerant charge is 1.7 kg with a CO2 equivalent of 2.4 t, which matters for leak checks and F-gas record keeping.
Triggers regulatory leak-check and documentation obligations.
“Refrigerant quantity 1,7 kg”
The unit weighs 476 kg and measures 938x1288x2003 mm, so floor loading and access routes must be verified.
Ensures the unit can be transported and positioned at the site.
“Overall dimensions WxDxH ** 938x1288x2003 mm”