The chiller delivers 18 kW cooling at 15 °C and 0 °C but only 6 kW at -20 °C, so sizing must be based on the lowest required operating temperature.
Cooling capacity collapses at low temperature; undersizing at -20 °C will cause loss of temperature control.
Source
“at 15°C 18 kW at 0°C 18 kW at -10°C 11 kW at -20°C 6 kW”
Selection
The unit uses R-449A with GWP 1397, which matters for refrigerant regulations and environmental compliance planning.
High-GWP refrigerant affects regulatory compliance, leak monitoring obligations and future serviceability.
Source
“R-449A (A1, H280) Global Warming Potential (GWP) 1397”
Installation
The chiller is air-cooled, so sufficient room ventilation and respect of space requirements are mandatory at the installation site.
Insufficient air flow around an air-cooled condenser reduces cooling capacity and can trip the refrigeration circuit.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Installation
The three-phase 400 V unit is supplied with cable but without a plug, so the customer must provide a suitable mains connection and protection.
Electrical installation planning must include a correctly rated plug, breaker and earthing by a qualified electrician.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Mains voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the unfavorable direction, so supply quality must be verified.
Out-of-tolerance supply can damage the compressor and electronics and void warranty claims.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
Cooling capacity data are valid at 20 °C room temperature and drop as ambient temperature rises, so hot plant rooms reduce performance.
Actual available cooling must be derated for the real ambient temperature at the installation site.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation
The permitted ambient temperature range is 5 to 40 °C, which defines where the unit may be installed and operated.
Operating outside this range risks compressor damage and invalidates performance specifications.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation
The pump delivers 210 l/min at zero head but only 77 l/min at 4.0 bar, so the system pressure drop must be checked against the pump curve.
External circuit resistance determines real flow; a high-pressure-drop loop may starve the process of flow.
Source
“max. delivery 210 l/min max. delivery pressure 4,7 bar”
Installation
The pump connection is G1 1/4 male, so hoses and adapters must match this thread size.
Wrong fittings cause leaks and delays during commissioning; correct couplings should be ordered in advance.
Source
“Pump connection G1 1/4 male”
Operation
Minimum filling capacity is 14 l and expansion volume is 39 l, which sets the heat transfer fluid charge and expansion allowance.
Underfilling causes pump cavitation and poor control; the expansion volume limits the total loop volume.
Source
“min. filling capacity 14 l Volume of expansion 39 l”
Safety
The unit has adjustable overtemperature protection per DIN 12876 and safety classification III/FL, relevant for flammable or sensitive media.
Overtemperature protection must be set correctly for the fluid used to prevent fire or fluid degradation.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Cumplimiento
The equipment is EMC class A group 1, so it is intended for industrial environments and may cause interference in residential areas.
Class A equipment may require additional filtering or separation from sensitive electronics in mixed environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit is IP20, so it must be installed in a dry, dust-free location protected from water ingress.
IP20 offers no water protection; washdown or outdoor installation is not permissible.
Source
“Degree of Protection IP20”
Selection
The unit is intended for externally closed applications, which constrains the hydraulic circuit design.
Open-loop or tank-vented circuits are not supported and would cause fluid loss and control problems.
Source
“For externally closed applications.”