Performance, temperature, and dimensional data per DIN 12876
p. 2table, 15 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
15 l
Volume of expansion
22 l
Overall dimensions WxDxH **
760x800x1615 mm
Net weight
401 kg
sound pressure level +/- 4 dB(A)
59 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current
24 A
Fuse (3 phase)
3x25A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S457845
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 180Tw-H (Order-No.: 3019.0044.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
18 kW
Cooling power at 0°C
18 kW
Cooling power at -10°C
13 kW
Cooling power at -20°C
6 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
2,6 kg
CO2 equivalent
3,6 t
Circulation pump
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
666 l/h
Consumption at water 15°C, flow -10°C
553 l/h
Consumption at water 15°C, flow -20°C
427 l/h
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires cooling water supply pressure between 1 bar minimum differential and 6 bar maximum, so the building water system must be sized within this window.
Incorrect cooling water pressure will prevent safe commissioning or damage the unit.
Source
“min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar”
Installation
The unit is water-cooled and must be fed with cooling water between 5°C and 20°C; below the condensation point the cooling water pipes inside the unit must be insulated.
Condensation on uninsulated pipes can cause water damage and electrical hazards.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling capacity is rated at 18 kW at 15°C and 0°C but drops to 13 kW at -10°C and 6 kW at -20°C, so low-temperature duty needs a larger unit or derating.
Sizing must account for the actual process temperature, not just the nominal 18 kW.
Source
“at 15°C
18 kW
at 0°C
18 kW
at -10°C
13 kW
at -20°C
6 kW”
Installation
The unit needs a 400 V 3-phase 50 Hz supply with a 3x25 A fuse and draws up to 24 A, so the electrical feed and protection must be sized accordingly.
Under-sized supply or fusing will trip or overheat the installation.
Source
“Power supply (3 Phase)
400V 3~ 50Hz
max. current
24 A
Fuse (3 phase)
3x25A”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the site plug.
Missing plug must be planned before commissioning to avoid downtime.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively but must not be combined in opposite directions, so the site supply quality must be verified.
Out-of-tolerance supply voids performance and may damage the compressor.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The refrigerant R-449A is classified A1/H280 with GWP 1397 and 2.6 kg charge (3.6 t CO2e), so leak detection and F-gas record keeping apply.
F-gas regulations and room ventilation requirements depend on charge and GWP.
Source
“R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
2,6 kg
CO2 equivalent
3,6 t”
Installation
The unit is IP20 and must be installed at ambient temperatures between 5°C and 40°C, so it cannot be placed in wet or hot plant rooms.
Wrong ambient conditions reduce cooling capacity and may void warranty.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The unit weighs 401 kg and measures 760x800x1615 mm, so floor loading and access routes must be checked before delivery.
Handling and rigging must be planned; the footprint and weight affect floor and door clearance.
Source
“Overall dimensions WxDxH **
760x800x1615 mm
Net weight
401 kg”
Operation
The pump delivers 210 l/min at zero head but only 31 l/min at 4.5 bar, so external circuit pressure drop must be matched to the pump curve.
A high-pressure-drop external loop will starve flow and reduce heat transfer.
Source
“max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 4,5 bar
31 l/min”
Operation
The expansion tank volume is 22 l and minimum filling capacity is 15 l, so the external loop volume plus unit must be checked against these limits.
Too small a system volume causes poor temperature stability; too large exceeds expansion capacity.
Source
“min. filling capacity
15 l
Volume of expansion
22 l”
Cumplimiento
The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.
Class A equipment may not be suitable for domestic or light-industrial locations without filtering.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality and treatment may be used.
Aggressive or untreated water will corrode or foul these materials and reduce heat transfer.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Selection
Cooling water consumption rises as process temperature falls: 666 l/h at 0°C, 553 l/h at -10°C and 427 l/h at -20°C at 15°C water inlet.
Facility cooling water capacity must cover the worst-case flow at the required process temperature.
Source
“Consumption at water 15°C, flow 0°C
666 l/h
Consumption at water 15°C, flow -10°C
553 l/h
Consumption at water 15°C, flow -20°C
427 l/h”