Contenido

  1. Overview p. 1
    Product name and general introduction to the Unichiller 180w
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 35 rows

Technical data according to DIN 12876 — Unichiller 180w (Order-No.: 3041.0001.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 20°C18 kW
at 0°C10 kW
at -10°C6 kW
at -20°C3,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,49 kg
CO2 equivalent2,1 t
Circulation pumpE1
max. delivery54 l/min
max. delivery pressure3,5 bar
Delivery at 0,3 bar52 l/min
Delivery at 0,5 bar48 l/min
Delivery at 1,0 bar42 l/min
Delivery at 1,5 bar36 l/min
Delivery at 2,0 bar28 l/min
Delivery at 2,5 bar21 l/min
Delivery at 3,0 bar11 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **940x1290x1130 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller is water-cooled, so a cooling water supply and return with G3/4 male connections must be provided at the installation site.

Determines the piping and utilities required before the unit can be commissioned.

Source
“Cooling water connection G3/4 male”
Installation

The unit requires a three-phase 400 V 50 Hz supply, so the electrical installation must be sized and protected accordingly.

Affects electrical design, cable sizing and whether a plug is supplied.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.

Prevents delays at commissioning because no plug is included in delivery.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Defines the acceptable cooling water supply conditions and insulation requirement.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.

The published 18 kW at 20 °C is only valid under the stated water conditions, so sizing must account for site water temperature.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

Only suitable cooling water may be used because the circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.

Water quality and treatment must be compatible with these materials to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

The site power quality must be checked against this combined tolerance before connection.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Cumplimiento

The equipment is EMC classified as Class A, Group 1, which has implications for use in residential or light-industrial environments.

Class A equipment may require additional EMC measures in sensitive or residential locations.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The refrigerant is R-449A with safety group A1 and GHS hazard H280, and the charge is 1.49 kg corresponding to 2.1 t CO2 equivalent.

Affects refrigerant handling, leak detection and F-gas reporting obligations.

Source
“R-449A (A1, H280)”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.

Restricts the installation environment and rules out washdown or outdoor siting.

Source
“Degree of Protection IP20”
Operation

The unit is intended for externally closed applications, which determines the permissible hydraulic circuit configuration.

The user circuit must be closed; open baths or unsealed loops are not permitted.

Source
“For externally closed applications.”
Installation

The minimum filling capacity is 18 l and the expansion volume is 48 l, which sets the minimum system volume and expansion allowance.

The external loop and expansion tank must be sized to meet these volume requirements.

Source
“min. filling capacity 18 l”
Installation

The overall dimensions are 940 x 1290 x 1130 mm and space requirements must be respected.

Confirms that clearance around the unit must be checked against the operating conditions before siting.

Source
“Please respect space requirements.”
Operation

The ambient temperature during operation must remain between 5 °C and 40 °C.

The room ventilation and climate must keep the installation within this ambient range.

Source
“max. ambient temperature 40 °C”