Unichiller 200T-H Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C20 kW
Cooling power at 0°C10 kW
Cooling power at -10°C5 kW
Cooling power at -20°C3 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x1582x1902 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller's cooling capacity drops as ambient temperature rises, so size the unit for the actual room conditions.

Cooling capacity ratings are referenced to 20 °C ambient; higher ambient reduces available capacity and may require a larger unit.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Three-phase units under 63 A are supplied with cable but no plug, so the installer must provide a suitable plug or hardwiring.

Affects electrical installation planning and bill of materials for the 400 V three-phase supply.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances must be met simultaneously, so verify the site supply before commissioning.

Prevents invalid supply combinations that could damage the compressor or pump.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit is IP20, so it must be installed in a dry indoor location protected from water ingress.

IP20 is not water-protected; placement in wet or washdown areas is not permitted.

Source
“Degree of Protection IP20”
Installation

The unit is air-cooled and requires adequate space for airflow, so respect the stated space requirements.

Insufficient clearance reduces cooling performance and can cause overheating.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Operation

The unit is intended only for externally closed applications, so do not use it with an open bath.

Using it in an open configuration risks overflow and incorrect operation.

Source
“For externally closed applications.”
Operation

The expansion tank is atmospheric open, so the installation must account for fluid level and venting.

An open expansion tank requires correct fluid filling and level monitoring.

Source
“Housing, atmospheric open expansion tank and copper soldered evaporator made of stainless steel.”
Safety

The refrigerant is R-449A with GHS H280, so handling and disposal must follow pressurized-gas safety rules.

H280 indicates a pressurized gas hazard; service personnel must be qualified and follow local regulations.

Source
“R-449A (A1, H280)”
Safety

The unit has adjustable overtemperature protection per DIN 12876, so set it to protect the application.

Proper setting of the overtemperature cut-out protects the process and the equipment.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance

The circulation pump delivers 210 l/min at zero head but only 31 l/min at 4.5 bar, so match the external circuit pressure drop.

External loop pressure drop determines actual flow; undersized piping will reduce cooling performance.

Source
“max. delivery 210 l/min”
Performance

Cooling capacity falls from 20 kW at 15 °C to 3 kW at -20 °C, so verify capacity at the actual operating temperature.

Selecting on the 15 °C rating alone will undersize the unit for low-temperature duty.

Source
“at 15°C 20 kW”
Cumplimiento

The unit is EMC Class A Group 1, so it may cause interference in residential environments.

Class A equipment is intended for industrial use and may require additional filtering in sensitive areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The refrigerant has a GWP of 1397, so leak checks and record-keeping are required under F-gas rules.

High GWP refrigerant triggers periodic leak inspection and documentation obligations.

Source
“Global Warming Potential (GWP) 1397”