Performance, temperature, and dimensional data per DIN 12876
p. 2table, 4 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/20
Technical data according to DIN 12876 — Unichiller 210Tw (Order-No.: 3020.0046.01)
Parameter
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 15°C
21 kW
Cooling power at 0°C
21 kW
Cooling power at -10°C
15,5 kW
Cooling power at -20°C
9,5 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
15 l
Volume of expansion
22 l
Overall dimensions WxDxH **
760x800x1615 mm
Power supply (3 Phase)
400V 3~ 50Hz
Pressure equipment category
I
** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1 bar and must not exceed 6 bar, so the site cooling water supply must be verified against these limits.
Insufficient differential pressure will prevent proper cooling; excessive pressure may damage the cooling circuit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is water-cooled and requires a cooling water connection of G3/4 male, so the site must provide a matching connection.
Mismatched connections require adapters and can delay commissioning.
Source
“Cooling water connection G3/4 male”
Operation
The cooling water inlet temperature must be between 5°C and 20°C; below the condensation point the cooling water pipes in the unit must be insulated.
Operating outside this range reduces cooling capacity or causes condensation and corrosion risk.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, so actual performance may differ on site.
Selection must account for site conditions; higher cooling water temperature reduces capacity and increases flow demand.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and - outlet.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Incompatible water chemistry can corrode or foul the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations (e.g., -5% voltage with +2% frequency) are not allowed.
Site power quality must be checked to avoid invalid supply conditions and equipment damage.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices with current consumption less than 63A are supplied with cable but without plug, so a suitable plug must be provided on site.
Electrical installation planning must include a compatible plug and connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The refrigerant is R-449A with GHS classification H280 (pressurized gas), so handling and ventilation precautions apply.
Personnel must be aware of pressurized gas hazards during installation and maintenance.
Source
“R-449A (A1, H280)”
Installation
The minimum filling capacity is 15 l and expansion volume is 22 l, so the external circuit volume and expansion vessel sizing must be considered.
Undersized or oversized external circuits affect temperature stability and pump performance.
Source
“min. filling capacity 15 l”
Operation
The unit has IP20 protection and must operate between 5°C and 40°C ambient temperature, so it must be installed in a dry, temperature-controlled environment.
IP20 offers no water protection; ambient limits affect reliability and warranty.
Source
“Degree of Protection IP20”
Installation
Space requirements must be respected as indicated in the operating conditions, so clearance around the unit must be planned.
Insufficient clearance can impede maintenance and cooling air flow.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Cumplimiento
The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance, so it is intended for industrial environments.
Class A equipment may cause interference in residential areas and requires appropriate EMC planning.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”