Installation
The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
Determines the required cooling water pump and piping design for installation.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is designed for cooling water intake temperatures between 5°C and 20°C, and pipes must be insulated if condensation occurs.
Ensures proper cooling water supply conditions and prevents condensation issues.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and flow rate may need to increase.
Affects sizing of cooling water supply and expected performance.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water should be used.
Guides water quality requirements to avoid corrosion or damage.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but certain combinations are not allowed.
Ensures power supply quality meets equipment requirements.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices with current consumption less than 63A are supplied with cable but without plug.
Affects electrical installation planning and plug procurement.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The refrigerant R-449A has a GHS classification of H280 (pressurized gas) and a GWP of 1397.
Important for safety handling and environmental compliance.
Source
“R-449A (A1, H280)”
Operation
The unit has a minimum filling capacity of 18 liters and an expansion volume of 48 liters.
Needed to prepare heat transfer fluid volume for operation.
Source
“min. filling capacity 18 l”
Installation
The unit requires a 400V 3-phase 50Hz power supply with max current 14A and fuse 3x16A.
Determines electrical supply requirements for installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Operation
The ambient temperature must be between 5°C and 40°C for proper operation.
Ensures the installation location meets environmental conditions.
Source
“min. ambient temperature 5 °C”
Cumplimiento
The equipment is classified as Class A, Group 1 for electromagnetic compatibility according to EN55011.
Important for EMC compliance in industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”