Unichiller 260T-H Technical Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data according to DIN 12876 including performance and dimensions
  3. p. 2 table, 3 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C26 kW
Cooling power at 0°C26 kW
Cooling power at -10°C16 kW
Cooling power at -20°C10 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x2172x1900 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller's cooling capacity drops as ambient temperature rises, so size the unit for the actual room conditions.

Prevents undersizing when the installation room is warmer than 20 °C.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Three-phase units below 63 A are supplied with cable but no plug, so the installer must provide a suitable plug.

Affects electrical installation planning and required site materials.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances must be met simultaneously, so verify the site supply before commissioning.

Ensures the 400 V 3~ supply is within the allowed combined tolerance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

Determines suitable installation environment and enclosure needs.

Source
“Degree of Protection IP20”
Installation

The pump connection is G1 1/4 male, so matching hose couplings must be provided for the external circuit.

Needed to order correct fittings and hoses for the process loop.

Source
“Pump connection G1 1/4 male”
Operation

The unit is intended only for externally closed applications, so an open process circuit is not permitted.

Prevents misuse with open tanks that could cause overflow or contamination.

Source
“For externally closed applications.”
Operation

Ambient temperature must stay between 5 °C and 40 °C for reliable operation.

Defines the allowable room conditions for warranty and performance.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so ventilation and safe handling procedures are required.

Informs site risk assessment and refrigerant handling requirements.

Source
“R-449A (A1, H280)”
Safety

The unit has adjustable overtemperature protection per DIN 12876, which must be set for the application.

Critical safety setting to protect the process and equipment from overheating.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance

Cooling capacity falls from 26 kW at 0 °C to 10 kW at -20 °C, so low-temperature duty needs careful sizing.

Prevents overestimating available capacity at the cold end of the range.

Source
“at -20°C 10 kW”
Performance

Pump delivery drops with rising pressure, from 202 l/min at 1.1 bar to 31 l/min at 4.5 bar.

Needed to check whether the external circuit's pressure drop is acceptable.

Source
“Delivery at 4,5 bar 31 l/min”
Cumplimiento

The equipment is Class A, Group 1 for EMC, so it may cause radio interference in residential environments.

Affects installation location and EMC precautions in sensitive areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Cumplimiento

The unit is compliant with US-SNAP for industrial process refrigeration, confirming its intended end-use.

Confirms regulatory suitability for industrial process cooling applications.

Source
“The US-SNAP end-use for this equipment is the industrial process refrigeration.”
Maintenance

The refrigerant has a GWP of 1397, so leak checks and refrigerant accounting are important.

Relevant for environmental compliance and service planning.

Source
“Global Warming Potential (GWP) 1397”