Product introduction and general description of the chiller
Specpp. 1-2
Technical data according to DIN 12876
p. 2table, 3 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/19
Technical data according to DIN 12876 — Unichiller 350T-H (Order-No.: 3021.0008.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,2 K
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
35 kW
Cooling power at 0°C
23 kW
Cooling power at -10°C
14 kW
Cooling power at -20°C
8 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump:
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
min. filling capacity
14 l
Volume of expansion
39 l
Overall dimensions WxDxH **
905x2172x1900 mm
Power supply (3 Phase)
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
** Overall dimensions WxDxH
Engineering Facts
12 facts · every claim linked to a page
Selection
The chiller's cooling capacity drops as ambient temperature rises, so selection must account for the actual room temperature at the installation site.
Cooling capacity is rated at 20°C ambient; higher ambient reduces available cooling and may require a larger unit.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so the installer must provide a suitable plug or hard-wired connection.
Determines what electrical termination hardware must be procured and planned for before commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances must be met simultaneously, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Ensures the site power quality is compatible with the compressor and electronics before installation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit requires respect for space requirements and operating conditions, which must be checked before siting.
Air-cooled units need adequate clearance and ambient conditions; ignoring this can derate or damage the chiller.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Operation
The chiller is intended only for externally closed applications, so the process loop must be closed.
Using it on an open loop would be outside the intended design and could cause overflow or contamination.
Source
“For externally closed applications.”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.
Informs safety planning for refrigerant leak scenarios and service access.
Source
“R-449A (A1, H280)”
Cumplimiento
The equipment emits Class A, Group 1 disturbances per EN55011, so it is intended for industrial electromagnetic environments.
Class A equipment may not be suitable for residential or light-commercial EMC environments without mitigation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation
The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.
Proper setting of the safety limiter protects the process and the chiller from overheating.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation
The pump connection is G1 1/4 male, and a matching hose coupling is included, so process piping must match this thread.
Determines the fittings and hoses needed for connecting the process loop.
Source
“Pump connection G1 1/4 male”
Selection
The minimum filling capacity is 14 l and expansion volume is 39 l, which sets the minimum heat-transfer fluid charge and expansion allowance.
Ensures the external loop and expansion vessel are sized so the pump and level monitoring work correctly.
Source
“min. filling capacity 14 l”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this range.
Operating outside the ambient range can reduce cooling capacity or trip protections.
Source
“max. ambient temperature 40 °C”
Selection
The chiller uses an air-cooled refrigerating machine that is CFC- and HCFC-free, which matters for environmental compliance and service.
Confirms refrigerant type and cooling method for site planning and environmental requirements.