Unichiller 350Tw Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller 350Tw chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Technical data according to DIN 12876 — Unichiller 350Tw (Order-No.: 3021.0010.01)

CharacteristicValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 15°C35 kW
Cooling power at 0°C25 kW
Cooling power at -10°C16 kW
Cooling power at -20°C10 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity15 l
Volume of expansion22 l
Overall dimensions WxDxH **760x800x1615 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI

** см. сноску в оригинале (Overall dimensions WxDxH **)

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar at the connection.

To ensure adequate cooling water flow and prevent damage to the refrigeration circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The cooling water inlet temperature must be between 5°C and 20°C; if below the condensation point, the cooling water pipes inside the unit must be insulated.

Operating outside this range can cause condensation, reduced cooling capacity, or damage.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

To properly size the cooling water supply and ensure the chiller meets process demands.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.

To avoid corrosion or fouling that could impair heat transfer and damage components.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but the combination of -5% voltage and +2% frequency is not allowed.

To ensure the power supply meets the allowed tolerances and avoid invalid combinations that could affect operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63A are supplied with a cable but without a plug.

To plan for proper electrical connection and obtain a suitable plug or hardwire the unit.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.

To assess electromagnetic compatibility and potential interference with sensitive equipment in the installation environment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The unit has a protection degree of IP20, so it must be installed in an environment protected from dust and water.

To ensure the installation location meets the environmental protection requirements.

Source
“Degree of Protection IP20”
Installation

The ambient temperature for operation must be between 5°C and 40°C.

To avoid operating outside the allowed ambient conditions, which can affect performance and reliability.

Source
“min. ambient temperature 5 °C”
Selection

The refrigerant is R-449A with a Global Warming Potential of 1397, which must be considered for environmental compliance and leak management.

To ensure compliance with regulations and to plan for proper handling and leak detection.

Source
“R-449A (A1, H280)”
Installation

The minimum filling capacity is 15 liters and the expansion volume is 22 liters, which must be considered when designing the external circuit.

To ensure the system has adequate fluid volume and expansion space for safe operation.

Source
“min. filling capacity 15 l”
Performance

The circulation pump delivers a maximum flow of 210 l/min and a maximum pressure of 4.7 bar, with flow decreasing at higher pressures.

To properly size the external circuit and ensure the pump can overcome system pressure losses.

Source
“max. delivery 210 l/min”