Product introduction and main features of the chiller
Specpp. 1-2
Technical data and performance specifications per DIN 12876
p. 2table, 8 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Overall dimensions WxDxH **
1000x1100x1635 mm
Net weight
634 kg
Power supply (3 Phase)
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/20
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Характеристика
Unichiller 500Tw-H
Order-No.
3030.0014.01
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
50 kW
Cooling power at 0°C
30 kW
Cooling power at -10°C
24 kW
Cooling power at -20°C
14 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump
D3
max. delivery
234 l/min
max. delivery pressure
4,9 bar
Delivery at 1,3 bar
197 l/min
Delivery at 1,5 bar
192 l/min
Delivery at 2,0 bar
178 l/min
Delivery at 2,5 bar
158 l/min
Delivery at 3,0 bar
137 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
86 l/min
Delivery at 4,5 bar
40 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1 1/4 male
Consumption at water 15°C, flow 15°C
2040 l/h
min. cooling water differential pressure
1,5 bar
max. cooling water pressure
6 bar
min. filling capacity
12,7 l
Volume of expansion
7 l
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires a three-phase 400 V supply and is delivered without a plug, so the site must provide a suitable 63 A-or-less three-phase connection and local isolation.
Determines electrical supply, cable termination and switchgear needed at installation.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Cooling water must be supplied at 5–20 °C inlet with at least 1.5 bar differential pressure and no more than 6 bar, so the facility water system must be checked before commissioning.
Cooling water conditions directly govern achievable cooling capacity and safe operation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water is colder than the ambient dew point, the cooling water pipes inside the unit must be insulated to prevent condensation.
Prevents condensation, corrosion and electrical risk in low-temperature cooling water installations.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Rated cooling capacity is only valid at 20 °C room temperature and 15 °C cooling water with 1.5 bar differential pressure, so higher water temperatures will reduce capacity and may require more flow.
Prevents overestimating cooling capacity when site conditions differ from the rating basis.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1,5 bar differential pressure between cooling water inlet and -outlet.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, brass, PA, PPE, PTFE and EPDM, so the facility water must be compatible with these materials.
Guides water quality and material compatibility checks to avoid corrosion or fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage and frequency tolerances are ±5 % and ±2 % respectively, but they may not be combined in opposite directions, so the site supply must be verified against this rule.
Defines acceptable power quality limits for warranty and reliable operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The unit is classified IP20, so it must be installed in a dry indoor location protected from water ingress and dust.
Restricts installation environment and enclosure protection requirements.
Source
“Degree of Protection
IP20”
Installation
Ambient temperature must stay between 5 °C and 40 °C, so the installation room needs adequate ventilation and heating.
Out-of-range ambient temperature affects performance, reliability and safety.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Selection
The refrigeration circuit uses R-449A with GWP 1397, which matters for refrigerant leakage reporting and environmental compliance.
Refrigerant type and GWP affect regulatory obligations and service planning.
Source
“R-449A (A1, H280)”
Selection
The pump delivers 234 l/min maximum at 4.9 bar, but flow falls to 40 l/min at 4.5 bar, so system pressure drop must be matched to the pump curve.
Ensures the external circuit resistance does not starve flow and reduce heat transfer.