Product introduction and general description of the chiller unit
Specpp. 1-2
Technical data, performance figures and operating parameters per DIN 12876
p. 2table, 10 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
min. filling capacity
12,7 l
Volume of expansion
7 l
Overall dimensions WxDxH **
1000x1100x1635 mm
Net weight
634 kg
Power supply (3 Phase)
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/21
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
70 kW
Cooling power at 0°C
50 kW
Cooling power at -10°C
30 kW
Cooling power at -20°C
20 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump
D3
max. delivery
234 l/min
max. delivery pressure
4,9 bar
Delivery at 1,3 bar
197 l/min
Delivery at 1,5 bar
192 l/min
Delivery at 2,0 bar
178 l/min
Delivery at 2,5 bar
158 l/min
Delivery at 3,0 bar
137 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
86 l/min
Delivery at 4,5 bar
40 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1 1/4 male
Consumption at water 15°C, flow 15°C
3780 l/h
Consumption at water 15°C, flow 0°C
2940 l/h
Consumption at water 15°C, flow -10°C
1980 l/h
min. cooling water differential pressure
1,5 bar
max. cooling water pressure
6 bar
Order-No.: 3032.0006.01; Unichiller 700Tw-H
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires cooling water inlet temperature between 5°C and 20°C, with insulation of cooling water pipes mandatory if below the condensation point.
Determines cooling water supply requirements and pipe insulation needs during installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
Cooling water differential pressure must be at least 1.5 bar and must not exceed 6 bar.
Defines the required cooling water supply pressure range for proper operation.
Source
“min. cooling water differential pressure 1,5 bar”
Installation
Three-phase devices with current consumption less than 63A are supplied with cable but without plug, so the customer must provide a suitable plug.
Affects electrical installation planning and procurement of the mating plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.
Ensures the site power supply stays within safe limits for the compressor and electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit has IP20 protection and must be installed in an environment with ambient temperature between 5°C and 40°C.
Restricts installation location to dry indoor areas with adequate ventilation and temperature control.
Source
“Degree of Protection IP20”
Selection
Cooling capacity drops as cooling water temperature rises, and cooling water flow rate may need to increase accordingly.
Affects sizing of the cooling water supply and expected performance at site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Selection
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality requirements to avoid corrosion and fouling of the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Safety
The refrigerant R-449A has safety group A1 and GHS hazard H280, and the unit has safety classification III / FL.
Informs risk assessment, ventilation requirements and refrigerant handling procedures.
Source
“R-449A (A1, H280)”
Safety
The unit includes adjustable overtemperature protection according to DIN 12876.
Provides a safety function that must be configured correctly for the application.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The unit is intended for externally closed applications and has an atmospheric open expansion tank.
Defines the permissible hydraulic circuit configuration and prevents misuse with open systems.
Source
“For externally closed applications.”
Operation
The pump connection and cooling water connection are both G1 1/4 male.
Specifies the thread type for hose couplings and piping adapters.
Source
“Pump connection G1 1/4 male”
Performance
Cooling power is 70 kW at 15°C, 50 kW at 0°C, 30 kW at -10°C and 20 kW at -20°C, valid at 20°C room temperature and 15°C cooling water inlet.
Allows correct sizing of the chiller for the required heat load at the target temperature.
Source
“at 15°C 70 kW”
Performance
Maximum pump delivery is 234 l/min and maximum delivery pressure is 4.9 bar, with delivery dropping at higher pressure.
Helps verify that the pump can overcome the pressure drop of the external circuit.
Source
“max. delivery 234 l/min”
Cumplimiento
The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.
Indicates EMC class and potential need for filtering or separation from sensitive equipment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”