The chiller requires a three-phase 400 V supply with a 3x16 A fuse, so the site must provide a suitable three-phase feed.
Determines electrical supply and protection sizing during installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
| Parameter | Value |
|---|---|
| Order-No. | 5502.0001.01 |
| Operating temperature range | -20...40 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Temperature stability at -10°C | 0,2 K |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Safety classification | I / NFL |
| Cooling power at 20°C | 10 kW |
| Refrigeration machine | water-cooled, natural refrigerant |
| Refrigerant (ASHRAE, GHS) | R-744 (A1, H280) |
| Global Warming Potential (GWP) | 1 |
| Refrigerant quantity | 2,1 kg |
| CO2 equivalent | 0,0 t |
| UN-number | UN 2857 |
| max. delivery | 50 l/min |
| max. delivery pressure | 3,5 bar |
| Pump connection | G1 1/4 male |
| Cooling water connection | G1/2 male |
| min. cooling water differential pressure | 1 bar |
| max. cooling water pressure | 6 bar |
| Volume of expansion | 10 l |
| Overall dimensions WxDxH ** | 550x760x1455 mm |
| Power supply (3 Phase) | 400V 3~ 50Hz |
| max. current (3 Phase) | 7 A |
| Fuse (3 phase) | 3x16A |
| Pressure equipment category | I |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No. | S611977 |
** Overall dimensions WxDxH
The chiller requires a three-phase 400 V supply with a 3x16 A fuse, so the site must provide a suitable three-phase feed.
Determines electrical supply and protection sizing during installation.
“Power supply (3 Phase) 400V 3~ 50Hz”
The unit is water-cooled and needs cooling water at 1–6 bar differential pressure with G1/2 male connections.
Cooling water supply pressure and connection must be verified before installation.
“min. cooling water differential pressure 1 bar”
Cooling water inlet must be kept between 5 °C and 20 °C, and pipes must be insulated if below the condensation point.
Operating outside this range risks condensation and reduced performance.
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Rated 10 kW cooling power applies only at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure.
Actual cooling capacity will differ if site conditions deviate from these reference values.
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Higher cooling water temperature reduces cooling power and may require increased cooling water flow.
Sizing of cooling water supply must account for derating at elevated water temperatures.
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.
Water chemistry must be compatible to avoid corrosion of these materials.
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Supply quality must be checked to avoid invalid voltage/frequency combinations.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Three-phase devices below 63 A are delivered with cable but without a plug, so a plug must be provided on site.
Electrical installation planning must include a suitable plug for the supplied cable.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
The unit uses R-744 refrigerant classified A1 with H280 hazard statement, requiring appropriate handling precautions.
Refrigerant safety classification affects handling and site safety requirements.
“R-744 (A1, H280)”
Ambient temperature must remain between 5 °C and 40 °C for safe operation.
Room conditions must be controlled to stay within the permitted ambient range.
“max. ambient temperature 40 °C”
The unit is intended for externally closed applications, which constrains how it can be integrated into the process.
Application type must match the closed-loop process design.
“For externally closed applications.”
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.
EMC class affects installation environment and potential interference considerations.
“Classification (disturbance) to EN55011: Class A, Group 1”