The chiller requires a three-phase 460 V 60 Hz supply, so the site must provide this voltage and frequency.
Determines electrical infrastructure and whether a transformer or different model is needed.
Source
“Power supply (3 Phase) 460V 3~ 60Hz”
| Parameter | Value |
|---|---|
| Order-No. | 5508.0004.01 |
| Operating temperature range | -20...40 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Temperature stability at -10°C | 0,2 K |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Safety classification | I / NFL |
| Cooling power at 20°C | 19 kW |
| Refrigeration machine | water-cooled, natural refrigerant |
| Refrigerant (ASHRAE, GHS) | R-744 (A1, H280) |
| Global Warming Potential (GWP) | 1 |
| UN-number | UN 2857 |
| max. delivery | 100 l/min |
| max. delivery pressure | 3,5 bar |
| Pump connection | G1 1/4 male |
| Cooling water connection | G3/4 male |
| min. cooling water differential pressure | 1 bar |
| max. cooling water pressure | 6 bar |
| Volume of expansion | 30 l |
| Overall dimensions WxDxH ** | 730x804x1738 mm |
| Power supply (3 Phase) | 460V 3~ 60Hz |
| Pressure equipment category | II |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No. | 1.0/26 |
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
The chiller requires a three-phase 460 V 60 Hz supply, so the site must provide this voltage and frequency.
Determines electrical infrastructure and whether a transformer or different model is needed.
“Power supply (3 Phase) 460V 3~ 60Hz”
The cooling water circuit needs a differential pressure of at least 1 bar and must not exceed 6 bar.
Ensures the water-cooled condenser receives adequate flow without overpressurizing the unit.
“min. cooling water differential pressure 1 bar”
Cooling water connections are G3/4 male, so mating hoses and fittings must match this thread.
Prevents delays and leaks during commissioning due to incompatible fittings.
“Cooling water connection G3/4 male”
The unit is designed for cooling water intake between 5 °C and 20 °C; below the condensation point the cooling water pipes inside the unit must be insulated.
Operating outside this range risks condensation, corrosion, or reduced performance.
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Rated cooling power of 19 kW applies only at 20 °C room temperature, 15 °C cooling water inlet, and 1 bar differential pressure.
Actual capacity will differ if site conditions deviate from these reference values.
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Prevents selecting a unit for a supply that falls outside the permitted combined tolerance.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
External branch circuit protection according to UL 489 is required.
Affects electrical panel design and compliance with North American installation codes.
“External branch circuit protection according to UL 489 required.”
Three-phase devices drawing less than 63 A are supplied with cable but without a plug.
The installer must provide a suitable plug or hard-wired connection.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
The refrigerant is R-744 (A1, H280), which is a non-flammable but high-pressure gas with an asphyxiation hazard in enclosed spaces.
Informs ventilation and safety measures for the machine room.
“R-744 (A1, H280)”
The chiller is water-cooled and uses natural refrigerant R-744 with GWP 1.
Requires a cooling water supply and may satisfy low-GWP refrigerant preferences.
“water-cooled, natural refrigerant”
The unit has IP20 protection, so it must be installed in a dry, indoor environment protected from water ingress.
Prevents placing the chiller where splashing or condensation could cause electrical faults.
“Degree of Protection IP20”
Ambient temperature must remain between 5 °C and 40 °C for safe operation.
Ensures the chiller is not installed in a location that exceeds its operating envelope.
“min. ambient temperature 5 °C”