Unichiller P010-H Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller P010-H chiller unit
  2. Spec p. 2
    Performance and technical data per DIN 12876
  3. p. 2 table, 11 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Overall dimensions WxDxH **350x496x622 mm
Net weight58 kg
sound pressure level +/- 4 dB(A)57 dB(A)
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current14 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S421582

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller P010-H (Order-No.: 3050.0004.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C
at 15°C1 kW
at 0°C0,8 kW
at -10°C0,5 kW
at -20°C0,15 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,33 kg
CO2 equivalent0,5 t
UN-numberUN 2857
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l

Cooling power at ambient temperature +20°C указана без значения (в тексте стоит точка).

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.

Defines the application window and excludes processes requiring lower or higher temperatures.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity falls sharply at low temperature, from 1 kW at 15 °C to only 0.15 kW at -20 °C, so the load must be checked at the coldest set point.

Prevents undersizing when the process requires deep cooling.

Source
“at -20°C 0,15 kW”
Installation

The unit requires a 220-240 V single- or two-phase supply with a maximum current of 14 A and a 16 A fuse.

Determines electrical supply, breaker rating and cabling before installation.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz max. current 14 A Fuse 16 A”
Installation

The chiller is air-cooled and must be installed where ambient temperature stays between 5 °C and 40 °C.

Ambient conditions affect cooling performance and reliable operation.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is specified at 20 °C room temperature and may drop if ambient temperature rises.

Prevents overestimating available cooling in warm plant rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Safety

The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, so refrigerant handling and leak response must follow safety rules.

Affects safety planning, ventilation and refrigerant leak procedures.

Source
“R-449A (A1, H280)”
Safety

The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used without a more resistant pump option.

Prevents chemical attack and premature pump failure in fluid selection.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection the manufacturer recommends mixing water with glysantine or ethylene glycol rather than using pure water.

Guides heat-transfer fluid selection for sub-zero operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The pump connection is G3/4 male, so hoses and fittings must match this thread.

Ensures correct plumbing connections during installation.

Source
“Pump connection G3/4 male”
Performance

The circulation pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with flow dropping to 6 l/min at 2.0 bar.

Allows checking whether the pump can overcome the pressure drop of the external circuit.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”
Installation

The unit needs at least 3.8 l filling capacity and has 1.7 l expansion volume, which sets the minimum fluid charge and expansion allowance.

Determines fluid volume and expansion vessel sizing for the loop.

Source
“min. filling capacity 3,8 l Volume of expansion 1,7 l”
Cumplimiento

The equipment is classified as pressure equipment category Art. 4.3 PED and has IP20 protection, so it must be installed in a dry indoor location.

Affects compliance documentation and environmental protection of the installation.

Source
“Pressure equipment category Art. 4.3 PED Degree of Protection IP20”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combining -5% voltage with +2% frequency is not allowed.

Prevents operation on unstable or mismatched power supplies.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Ensures the correct plug is ordered for the installation site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”