Unichiller P012-H Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller P012-H chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. Spec p. 2
    Detailed parameter table according to DIN 12876
  4. p. 2 table, 10 rows
  5. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Net weight66 kg
sound pressure level +/- 4 dB(A)59 dB(A)
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current15 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S522852

Technical data according to DIN 12876 — Unichiller P012-H (Order-No.: 3009.0196.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C
at 15°C1,2 kW
at 0°C1 kW
at -10°C0,7 kW
at -20°C0,25 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,45 kg
CO2 equivalent0,6 t
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **420x487x579 mm

** см. сноску в оригинале (в тексте страницы расшифровка не приведена).

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window.

Defines whether the unit can cover the required process temperatures at all.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity falls sharply at low temperatures, from 1.2 kW at +15 °C ambient to only 0.25 kW at -20 °C, so the heat load must be checked at the coldest operating point.

Sizing on nominal capacity alone will undersize the chiller for low-temperature duty.

Source
“at -20°C 0,25 kW”
Installation

The unit requires a 220-240 V single/two-phase 50/60 Hz supply with a 16 A fuse and up to 15 A current draw.

Determines the electrical feed, protection device and plug/cable configuration at the installation site.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz max. current 15 A Fuse 16 A”
Installation

Voltage and frequency tolerances are limited to ±5 % and ±2 % simultaneously, and certain combinations are explicitly not allowed.

Site power quality must be verified before commissioning to avoid invalid supply conditions.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The chiller is air-cooled, so ambient temperature must stay between 5 °C and 40 °C and space requirements around the unit must be respected.

Air-cooled units need adequate room ventilation and clearance or cooling capacity will drop.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity data are valid at 20 °C room temperature and may drop if ambient temperature rises.

Selection calculations must account for the actual ambient conditions in the installation room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic must be ordered.

Prevents chemical attack and premature pump failure when choosing the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

Guides correct heat transfer fluid preparation for sub-zero operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.

The correct plug type must be stated on the order to avoid rework at installation.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Confirms built-in safety cut-out for unattended operation and risk assessment.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Cumplimiento

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require EMC precautions in residential or sensitive laboratory environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The refrigerant is R-449A with GWP 1397 and a charge of 0.45 kg, giving 0.6 t CO2 equivalent.

Relevant for refrigerant handling, leak checks and environmental reporting obligations.

Source
“Refrigerant quantity 0,45 kg CO2 equivalent 0,6 t”
Installation

The pump connection is G3/4 male and the minimum filling capacity is 3.8 l with 1.7 l expansion volume.

Needed to select hoses, couplings and the correct heat transfer fluid fill volume.

Source
“Pump connection G3/4 male min. filling capacity 3,8 l Volume of expansion 1,7 l”
Performance

The circulation pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with flow falling as pressure rises.

Must be checked against the pressure drop of the external circuit to ensure adequate flow.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”