Unichiller P012 Recirculating Chiller Datasheet

Contenido

  1. Overview p. 1
    Product introduction and model identification for Unichiller P012
  2. Spec p. 2
    Performance and technical specifications according to DIN 12876
  3. p. 1 table, 32 rows

Unichiller P012 — Technical data according to DIN 12876

ParameterValue
Order-No.3009.0161.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C1,2 kW
at 0°C1 kW
at -10°C0,7 kW
at -20°C0,25 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,35 kg
CO2 equivalent0,5 t
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight65 kg
sound pressure level +/- 4 dB(A)59 dB(A)
Power supply requirement208V 2~ 60Hz

** — сноска в исходном тексте не раскрыта.

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so the process must stay within these limits.

Defines the application window and excludes processes requiring lower or higher temperatures.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity drops sharply at low temperatures, from 1.2 kW at +15 °C to only 0.25 kW at -20 °C, so sizing must use the capacity at the actual set point.

Prevents undersizing when the process runs cold; capacity at -20 °C is roughly one-fifth of that at +15 °C.

Source
“at 15°C 1,2 kW at 0°C 1 kW at -10°C 0,7 kW at -20°C 0,25 kW”
Performance

Cooling capacity is specified at 20 °C room temperature and may fall if ambient rises, so the installation room must be kept cool.

Air-cooled units lose capacity in warm rooms; ambient conditions must be considered during selection and site planning.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 208 V two-phase 60 Hz supply with a 7 A max current and 16 A fuse, which must be provided at the installation site.

Determines the electrical supply, cabling and protection device needed before installation.

Source
“Power supply requirement 208V 2~ 60Hz”
Installation

Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the unfavorable direction, so supply quality must be checked.

Avoids operating the compressor outside allowed supply conditions, which could damage it or void warranty.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are delivered with cable but without plug, so a suitable plug must be sourced and fitted on site.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant R-449A is classified A1 with hazard statement H280 (pressurized gas), so handling and ventilation precautions apply.

Informs safety assessment, labeling and room ventilation requirements for the refrigerant charge.

Source
“R-449A (A1, H280)”
Cumplimiento

The refrigerant has a GWP of 1397 and a CO2 equivalent of 0.5 t, which matters for F-gas reporting and leak-check obligations.

Determines regulatory obligations for refrigerant containment, leak checks and documentation.

Source
“Global Warming Potential (GWP) 1397”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must not be used as heat transfer fluid.

Prevents chemical attack and pump failure; an alternative resistant plastic is only available on request.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Avoids freezing and damage when operating near or below 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The pump connection is G3/4 male and hose couplings are included, so mating hoses and fittings must match this thread.

Ensures correct plumbing interface and avoids adapters or leaks at commissioning.

Source
“Pump connection G3/4 male”
Installation

The unit needs a minimum filling capacity of 3.8 l plus 1.7 l expansion volume, so the external loop volume must be planned accordingly.

Determines heat transfer fluid quantity and expansion vessel sizing for the installation.

Source
“min. filling capacity 3,8 l”
Installation

The unit is IP20 and must be installed where ambient temperature stays between 5 and 40 °C.

IP20 offers no water protection and the ambient limits define acceptable site conditions.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Maintenance

The unit emits 59 dB(A) sound pressure, which must be considered for workplace noise and room placement.

Affects installation location and possible hearing protection or acoustic measures.

Source
“sound pressure level +/- 4 dB(A) 59 dB(A)”