Performance, temperature, and dimensional data per DIN 12876
p. 2table, 9 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Overall dimensions WxDxH **
350x496x622 mm
Power supply requirement
220-240V 1~/2~ 50/60Hz
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.1/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller P012w-H (Order-No. 3009.0239.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at ambient temperature +20°C at 15°C
1,2 kW
Cooling power at ambient temperature +20°C at 0°C
1 kW
Cooling power at ambient temperature +20°C at -10°C
0,7 kW
Cooling power at ambient temperature +20°C at -20°C
0,25 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
60 l/h
Consumption at water 15°C, flow 0°C
57 l/h
Consumption at water 15°C, flow -10°C
48 l/h
min. cooling water differential pressure
1 bar
Cooling power values are sub-rows under 'Cooling power at ambient temperature +20°C'.
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller requires cooling water with a minimum differential pressure of 1 bar between inlet and outlet to operate correctly.
Ensures the water-cooled refrigeration unit receives adequate flow; insufficient differential pressure will impair cooling performance.
Source
“min. cooling water differential pressure 1 bar”
Installation
The cooling water intake temperature must be between 5°C and 20°C; if it falls below the condensation point, the cooling water pipes in the unit must be insulated.
Defines the permissible cooling water supply conditions and mandates insulation to prevent condensation when water is colder than the dew point.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Engineers must account for reduced cooling capacity and higher water consumption when the available cooling water is warmer than the rated 15°C.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The maximum permissible cooling water pressure is 6 bar.
Prevents overpressure damage to the cooling water circuit; supply pressure must be limited accordingly.
Source
“max. cooling water pressure 6 bar”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must not be used; water with glysantine or ethylene glycol is recommended for freeze protection.
Prevents chemical attack and premature failure of the pump; dictates compatible heat transfer fluids.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.
Defines the acceptable power supply envelope; violating it may cause malfunction or damage and affects site power quality requirements.
Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento
The unit is classified as EN55011 Class A, Group 1 for electromagnetic disturbance.
Class A equipment is intended for industrial environments; informs EMC planning and restricts use in residential areas without additional measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit has IP20 degree of protection, so it must be installed in a dry, protected environment.
IP20 offers no water protection; installation location must avoid moisture and dust ingress.
Source
“Degree of Protection IP20”
Installation
The ambient temperature at the installation site must be between 5°C and 40°C.
Operating outside this range may affect cooling performance, reliability, and warranty; constrains room conditioning requirements.
Source
“min. ambient temperature 5 °C”
Selection
The refrigerant R-449A has a Global Warming Potential of 1397, which is relevant for environmental compliance and leak reporting.
GWP affects regulatory obligations (e.g., F-gas reporting, leak checks) and environmental impact assessment during equipment selection.
Source
“Global Warming Potential (GWP) 1397”
Installation
The pump connection is G3/4 male and the cooling water connection is G1/2 male, determining the mating fittings required.
Ensures correct hose couplings and adapters are procured; mismatched threads prevent commissioning.
Source
“Pump connection G3/4 male”
Operation
The minimum filling capacity is 3.8 l and the expansion volume is 1.7 l, which sets the minimum heat transfer fluid charge and expansion allowance.
Prevents underfilling and overflow; needed to size the fluid charge and expansion vessel correctly.
Source
“min. filling capacity 3,8 l”
Selection
The unit is water-cooled and uses a CFC- and HCFC-free refrigerant, which affects site cooling water availability and environmental compliance.
Requires a cooling water supply at the installation site and supports compliance with refrigerant regulations.