Technical data according to DIN 12876 — Unichiller P012w (Order-No.: 3009.0231.01)
Parameter
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C
at 15°C
1,2 kW
at 0°C
1 kW
at -10°C
0,7 kW
at -20°C
0,25 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,19 kg
CO2 equivalent
0,3 t
Circulation pump
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
60 l/h
Consumption at water 15°C, flow 0°C
57 l/h
Consumption at water 15°C, flow -10°C
48 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
Строка «Cooling power at ambient temperature +20°C» напечатана как заголовок группы; значения приведены для подпунктов at 15°C, at 0°C, at -10°C, at -20°C.
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
To ensure the cooling water circuit can provide adequate flow and pressure without overpressurizing the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The cooling water intake temperature must be between 5°C and 20°C; if it falls below the condensation point, the cooling water pipes in the unit must be insulated.
To avoid condensation and ensure proper operation of the refrigeration system.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature increases, and the required cooling water flow rate may increase.
To properly size the cooling water supply and predict performance under actual site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation
The plastic pump is not resistant to acetone and polyglycols; for freeze protection, use water mixed with glysantine or ethylene glycol instead.
To select a compatible heat transfer fluid and avoid damaging the pump.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation
The unit requires a power supply of 220-240V 1~/2~ 50/60Hz with a maximum current of 6 A and a fuse rated between 10 A and 16 A.
To ensure the electrical supply and protection devices match the unit's requirements.
Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation
Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but combinations such as -5% voltage and +2% frequency are not allowed.
To verify that the site power quality is within acceptable limits for safe operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit has a degree of protection IP20, so it must be installed in a dry environment protected from water ingress.
To ensure the installation location meets the environmental protection requirements.
Source
“Degree of Protection IP20”
Installation
The ambient temperature must be between 5°C and 40°C for proper operation.
To avoid operating the chiller outside its specified environmental limits.
Source
“min. ambient temperature 5 °C”
Installation
The pump connection is G3/4 male and the cooling water connection is G1/2 male, so matching fittings must be provided.
To prepare correct plumbing adapters and hoses before installation.
Source
“Pump connection G3/4 male”
Safety
The refrigerant R-449A has safety group A1 and GHS hazard H280, and the unit contains 0.19 kg with a CO2 equivalent of 0.3 t.
To assess refrigerant safety classification and environmental reporting requirements.
Source
“R-449A (A1, H280)”
Cumplimiento
The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.
To ensure electromagnetic compatibility with other equipment in the installation environment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Three-phase devices with current consumption less than 63A are supplied with a cable but without a plug.
To plan for providing a suitable plug and electrical connection on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”