Product introduction and main features of Unichiller P015
Specpp. 1-2
Technical data, performance and dimensions per DIN 12876
p. 2table, 7 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
max. current
8,5 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S609077
Unichiller P015 — Technical data according to DIN 12876
Parameter
Value
Order-No.
3051.0046.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C
at 15°C
1,5 kW
at 0°C
1 kW
at -10°C
0,7 kW
at -20°C
0,3 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,16 kg
CO2 equivalent
0,0 t
Circulation pump:
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Overall dimensions WxDxH **
420x487x579 mm
Net weight
61 kg
sound pressure level +/- 4 dB(A)
63 dB(A)
Power supply requirement
220-240V 1~/2~ 50Hz
** Overall dimensions WxDxH
Engineering Facts
13 facts · every claim linked to a page
Selection
The chiller uses R-290 (A3, H220) refrigerant, which is flammable, so the installation location must be assessed for A3 refrigerant safety requirements.
A3 flammability classification affects room ventilation, ignition-source control, and local refrigerant regulations during selection and installation.
Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Selection
Cooling capacity drops from 1.5 kW at +15 °C to 0.3 kW at -20 °C, so the required load must be checked at the actual operating temperature.
Sizing only on nominal capacity will undersize the unit at low temperatures and cause insufficient cooling.
Source
“at 15°C 1,5 kW at 0°C 1 kW at -10°C 0,7 kW at -20°C 0,3 kW”
Performance
Cooling capacity is rated at 20 °C room temperature and may fall if ambient temperature rises, so the installation room must stay within the ambient limits.
High ambient temperature reduces available cooling and can make the chiller unable to hold the set point.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit requires 220-240 V single/two-phase 50 Hz supply with 8.5 A max current and a 16 A fuse, so the electrical supply and protection must be sized accordingly.
Correct breaker/fuse rating and supply configuration are needed to avoid nuisance trips and ensure safe operation.
Source
“max. current 8,5 A Fuse 16 A”
Installation
Voltage and frequency tolerances are ±5% and ±2% but must not be combined in the unfavorable direction, so supply quality must be verified before commissioning.
Out-of-tolerance supply can damage the compressor and pump and void warranty.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The pump connection is G3/4 male and the minimum filling capacity is 3.8 l, so hoses and expansion volume must be planned for this circuit.
Mismatched fittings or insufficient fill volume will prevent proper commissioning and circulation.
Source
“Pump connection G3/4 male min. filling capacity 3,8 l”
Operation
The plastic pump is not resistant to acetone and polyglycols, so these fluids must not be used without a more resistant pump option.
Using incompatible fluid will damage the pump and contaminate the loop.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Prevents freezing and damage when operating near or below 0 °C.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The unit is IP20 and must be installed in an environment between 5 °C and 40 °C ambient temperature.
IP20 offers no water protection and ambient limits define where the chiller can be safely located.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Cumplimiento
The equipment is EMC Class A Group 1, so it may cause radio interference in residential areas and is intended for industrial environments.
Class A equipment may not be suitable for residential use without additional EMC measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are supplied with a country-specific plug that must be specified at order, while three-phase devices below 63 A come without a plug.
Missing plug specification at order leads to delays and additional electrical work on site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Maintenance
The refrigerant charge is only 0.16 kg with GWP 0.02, so leak-check and service procedures must account for a small A3 charge.
Small flammable charge affects service safety and refrigerant record-keeping obligations.
Source
“Refrigerant quantity 0,16 kg CO2 equivalent 0,0 t”
Performance
The pump delivers 25 l/min max and 6 l/min at 2.0 bar, so system pressure drop must be checked against the pump curve.
High-pressure-drop loops may need the P model's higher discharge pressure or a different pump selection.