Selection
The chiller is rated for a -10 to 100 °C operating range, so it suits both sub-ambient cooling and moderate heating duties.
Defines whether the unit can cover the required process temperature window before specifying it.
Source
“Operating temperature range -10...100 °C”
Performance
Cooling capacity drops with lower temperature (2.2 kW at 15 °C, 1.6 kW at 0 °C, 1 kW at -10 °C), so size the load against the actual set point.
Prevents undersizing when the application runs cold rather than near ambient.
Source
“at 15°C 2,2 kW at 0°C 1,6 kW at -10°C 1 kW”
Installation
The unit requires a 208-230 V single/two-phase 60 Hz supply with 15 A max current and a 16 A fuse.
Determines electrical feed, breaker sizing and whether a 60 Hz site is compatible.
Source
“Power supply requirement 208-230V 1~/2~ 60Hz max. current 15 A Fuse 16 A”
Installation
Voltage and frequency tolerances must be met simultaneously (±5% V with ±2% f), and a -5% voltage with +2% frequency combination is not allowed.
Guides site power quality checks to avoid invalid supply conditions and warranty issues.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Ambient temperature must stay between 5 and 40 °C, and rising ambient temperature may reduce cooling capacity.
Affects room ventilation/placement planning and realistic capacity expectations.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Safety
The refrigerant is R-449A (A1, H280) with GWP 1397 and 0.56 kg charge, giving 0.8 t CO2 equivalent.
Needed for refrigerant handling, leak reporting and environmental compliance planning.
Source
“R-449A (A1, H280)”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump specified.
Prevents chemical attack on the pump when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
Ensures correct coolant preparation for sub-zero operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
Avoids a missing or wrong plug on delivery and rework at commissioning.
Source
“with power cable and country-specific plug (please specify when ordering)”
Cumplimiento
The unit is EMC classified as Class A, Group 1 to EN55011, which has implications for use in residential environments.
Class A equipment may not be suitable for domestic/low-voltage residential supply without mitigation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit has IP20 protection, so it must be installed in a dry, protected location away from water ingress.
Restricts placement options and enclosure requirements at the installation site.
Source
“Degree of Protection IP20”
Maintenance
The expansion vessel has a 1.7 l volume and a minimum filling capacity of 3.8 l applies to the system.
Sets the minimum coolant charge and expansion headroom for loop design.
Source
“min. filling capacity 3,8 l expansion tank 1,7 l”
Performance
The pump delivers 21 l/min at 0.5 bar down to 6 l/min at 2.0 bar, with 25 l/min max delivery and 2.5 bar max pressure.
Allows checking whether the pump curve covers the pressure drop of the external circuit.
Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”
Safety
The unit includes adjustable overtemperature protection according to DIN 12876 and safety classification I/NFL.
Confirms built-in thermal safety for unattended or sensitive processes.
Source
“With adjustable overtemperature protection according to DIN 12876.”