Selection
The chiller is rated for an operating temperature range of -10 to 40 °C, so process requirements outside this window cannot be met.
Defines the process temperature envelope for which the unit can be selected.
Source
“Operating temperature range
-10...40 °C”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects site safety classification and handling requirements.
Flammable refrigerant classification drives installation ventilation and safety measures.
Source
“R-290 (A3, H220)”
Installation
The unit requires a 220-240 V single-phase 50 Hz supply with a minimum 10 A fuse, so the electrical circuit must be sized accordingly.
Determines the electrical supply and protection device needed at the installation site.
Source
“Power supply requirement
220-240V 1~ 50Hz”
Installation
The pump connection is G3/4 male, so mating hoses and couplings must match this thread.
Ensures correct hydraulic connection hardware is ordered and installed.
Source
“Pump connection
G3/4 male”
Performance
Cooling capacity drops from 2.2 kW at 15 °C to 1 kW at -10 °C, so capacity must be checked at the actual operating temperature.
Prevents oversizing or undersizing based on a single nominal capacity figure.
Source
“at 15°C
2,2 kW
at 0°C
1,6 kW
at -10°C
1 kW”
Performance
The circulation pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar, so system pressure drop must be matched to the pump curve.
Flow at the actual system backpressure determines whether the pump is adequate.
Source
“at 0,5 bar
21 l/min
at 1,0 bar
17 l/min
at 1,5 bar
11 l/min
at 2,0 bar
6 l/min”
Installation
The minimum filling capacity is 3.8 l and the expansion tank is 1.7 l, so the loop volume and expansion allowance must be planned.
Ensures the circuit has sufficient fluid volume and expansion capacity.
Source
“min. filling capacity
3,8 l
expansion tank
1,7 l”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.
Restricts installation environment and prevents moisture-related failures.
Source
“Degree of Protection
IP20”
Operation
The minimum ambient temperature is 5 °C and the maximum is 40 °C, so the installation room must stay within this range.
Ambient conditions outside this range affect reliable operation and cooling performance.
Source
“min. ambient temperature
5 °C”
Performance
Cooling capacity data are valid at 20 °C room temperature and may drop if ambient temperature rises, so site conditions must be considered.
Prevents expecting nominal capacity in a warmer installation room.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Defines acceptable supply quality for warranty and reliable operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The plastic pump is not resistant to acetone and polyglycols, so these fluids must be avoided or a more resistant pump specified.
Prevents chemical attack and premature pump failure from incompatible heat transfer fluids.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Guides correct heat transfer fluid preparation to avoid freezing damage.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.
Ensures the correct plug type is delivered for the installation country.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”