Product introduction and general description of the Unichiller P025-H OLÉ
Specpp. 1-2
Technical data, performance figures and operating parameters per DIN 12876
p. 2table, 3 rows
p. 1table, 39 rows
Technical data according to DIN 12876
Parameter
Value
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S612190
Unichiller P025-H OLÉ — Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-10...100 °C
temperature set point / display
digital
Internal temperature sensor
Pt100
Resolution of display
0,1 K
Interface digital
USB (Device), RS232
Interface
Temperature stability at -10°C
0,2 K
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at 15°C
2,5 kW
Cooling power at 0°C
2 kW
Cooling power at -10°C
1,2 kW
Refrigeration machine
air-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,23 kg
CO2 equivalent
0,0 t
Circulation pump
B
at 0,5 bar
21 l/min
at 1,0 bar
17 l/min
at 1,5 bar
11 l/min
at 2,0 bar
6 l/min
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Pump connection
G3/4 male
min. filling capacity
3,8 l
expansion tank
1,7 l
Overall dimensions WxDxH **
460x590x743 mm
Net weight
81 kg
sound pressure level +/- 4 dB(A)
64 dB(A)
Power supply requirement
220-240V 1~ 50Hz
max. current
16 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller is rated for an operating temperature range of -10 to 100 °C, so it can serve both sub-ambient cooling and moderate heating tasks.
Defines whether the unit can cover the required process temperature window before any other sizing is done.
Source
“Operating temperature range
-10...100 °C”
Selection
Cooling capacity falls from 2.5 kW at 15 °C to 1.2 kW at -10 °C, so the unit must be sized against the actual operating point, not the nominal figure.
Prevents oversizing assumptions at low temperature where available cooling is roughly halved.
Source
“Cooling power
at 15°C
2,5 kW
at 0°C
2 kW
at -10°C
1,2 kW”
Safety
The refrigerant is R-290 (propane, A3 flammability class, H220), which imposes flammable-refrigerant handling and ventilation requirements on the installation.
A3 refrigerants require ignition-source control and room ventilation planning, and affect service and compliance approvals.
Source
“R-290 (A3, H220)”
Installation
The unit draws up to 16 A on a 220–240 V single-phase supply and is protected by a 16 A fuse, so the site circuit must be dedicated and adequately rated.
Determines cable cross-section, breaker rating and whether existing site power is sufficient.
Source
“Power supply requirement
220-240V 1~ 50Hz
max. current
16 A
Fuse
16 A”
Installation
The enclosure is rated IP20, so the chiller must be kept in a dry, clean location and cannot be washed down or exposed to splashing.
Constrains placement in the lab or plant and rules out wet or washdown environments.
Source
“Degree of Protection
IP20”
Installation
The pump connection is G3/4 male, so hoses or adapters must match this thread size at the circuit interface.
Avoids ordering incompatible fittings and delays during commissioning.
Source
“Pump connection
G3/4 male”
Performance
Pump delivery drops from 21 l/min at 0.5 bar to 6 l/min at 2.0 bar, so high-pressure-drop loops must be checked against the actual pump curve.
Ensures the required flow is still achievable in a restrictive external circuit.
Source
“at 0,5 bar
21 l/min
at 1,0 bar
17 l/min
at 1,5 bar
11 l/min
at 2,0 bar
6 l/min”
Operation
Cooling capacity is specified at 20 °C room temperature and may drop as ambient rises, so the installation room must stay within the allowed ambient window.
Prevents underperformance when the chiller is placed in a warm or poorly ventilated room.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The permitted ambient temperature range is 5 to 40 °C, which sets the boundary for the room where the air-cooled unit is installed.
Defines site ventilation and climate requirements before installation.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump specified.
Prevents chemical attack and premature pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection the manufacturer recommends water mixed with glysantine or ethylene glycol rather than pure water.
Guides the correct heat transfer fluid choice for sub-zero operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Voltage and frequency tolerances are coupled: ±5 % voltage is only allowed with ±2 % frequency, and the combination -5 % voltage with +2 % frequency is not permitted.
Prevents operation on unstable or generator-supplied mains that could damage the compressor.
Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento
The unit is EMC-classified as Class A, Group 1 to EN55011, so it is intended for industrial environments rather than residential ones.
Affects where the chiller may be installed and what EMC measures are needed nearby.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single- and two-phase units are delivered with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct plug type is supplied and avoids rework at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”