The chiller is rated for an operating temperature range of -10 to 40 °C, so it cannot be used for sub-zero process cooling below -10 °C.
Defines the process temperature window the unit can serve and excludes colder applications.
Source
“Operating temperature range
-10...40 °C”
Performance
Cooling capacity drops from 2.5 kW at 15 °C to 1.2 kW at -10 °C, so the unit must be sized against the actual process temperature, not the nominal maximum.
Prevents undersizing when the process runs near the low end of the temperature range.
Source
“Cooling power
at 15°C
2,5 kW
at 0°C
2 kW
at -10°C
1,2 kW”
Installation
The unit requires a 220-240 V single-phase 50 Hz supply with a 16 A fuse and draws up to 10.5 A, so the site electrical supply must be sized accordingly.
Determines the electrical connection, breaker/fuse rating and cable sizing at the installation site.
Source
“Power supply requirement
220-240V 1~ 50Hz
max. current
10,5 A
Fuse
16 A”
Installation
The chiller is air-cooled and must be installed where ambient temperature stays between 5 and 40 °C, with space requirements respected.
Ambient conditions affect cooling capacity and reliable operation; the room must be within this range.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
Cooling capacity data are valid at 20 °C room temperature and may drop if ambient temperature rises, so capacity must be derated for warmer rooms.
Prevents overestimating available cooling in a warm plant room.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the combination of -5 % voltage with +2 % frequency is not allowed.
Defines acceptable supply quality and a prohibited tolerance combination for the electrical installation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The plastic pump is not resistant to acetone and polyglycols, so these media must not be used in the circuit.
Prevents chemical attack on the pump and premature failure when selecting the heat transfer fluid.
Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
Guides the choice of heat transfer fluid to avoid freezing at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct plug type is ordered for the destination country.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Cumplimiento
The equipment is classified as EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments, not residential ones.
Affects EMC compliance and suitability of the installation location.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The refrigerant is R-449A with GHS classification A1/H280 and a GWP of 1397, so refrigerant handling and leak reporting obligations apply.
Informs refrigerant safety classification and environmental obligations for the site.
Source
“R-449A (A1, H280)”
Maintenance
The refrigerant charge is 0.5 kg, corresponding to 0.7 t CO2 equivalent, which is relevant for leak-check and F-gas record keeping.
Determines whether periodic leak checks and CO2-equivalent records are required.
Source
“Refrigerant quantity
0,5 kg
CO2 equivalent
0,7 t”
Installation
The pump connection is G3/4 male and the minimum filling capacity is 3.8 l, so the external circuit and fluid volume must be planned accordingly.
Defines the hydraulic connection size and minimum system fill volume for installation.
Source
“Pump connection
G3/4 male
min. filling capacity
3,8 l”
Performance
The circulation pump delivers 25 l/min maximum at up to 2.5 bar, with flow falling as pressure rises, so system pressure drop must be checked against the pump curve.
Ensures the pump can overcome the pressure drop of the external circuit.