Unichiller P050 Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 10 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Net weight268 kg
sound pressure level +/- 4 dB(A)62 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)8 A
Fuse (3 phase)3x10A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S318109

Unichiller P050 Technical data according to DIN 12876

CharacteristicValue
Order-No.3038.0004.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 20°C5 kW
at 0°C3,4 kW
at -10°C2,1 kW
at -20°C0,7 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,03 kg
CO2 equivalent1,4 t
Circulation pump:E2
max. delivery130 l/min
max. delivery pressure5,7 bar
Delivery at 2,4 bar116 l/min
Delivery at 2,5 bar115 l/min
Delivery at 3,0 bar108 l/min
Delivery at 3,5 bar100 l/min
Delivery at 4,0 bar92 l/min
Delivery at 4,5 bar80 l/min
Delivery at 5,0 bar59 l/min
Delivery at 5,5 bar20 l/min
Pump connectionG1 1/4 male
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1165 mm

** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is intended only for externally closed applications, so an open bath or vented loop is not permitted.

Defines the hydraulic circuit type allowed and prevents selecting the unit for open-tank duty.

Source
“For externally closed applications.”
Installation

The unit requires a three-phase 400 V 50 Hz supply with 8 A max current and 3x10 A fusing.

Determines site power availability, cable sizing and breaker rating before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the site plug.

Affects electrical installation scope and ordering of the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances must be met simultaneously; a -5% voltage with +2% frequency is not allowed.

Prevents connecting the chiller to a supply outside the permitted combined tolerance window.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit must be installed where ambient temperature stays between 5 °C and 40 °C.

Room conditions outside this range invalidate performance and may damage the refrigeration circuit.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and drops if ambient temperature rises.

Selection must account for actual plant room temperature, not just the nominal 5 kW.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Selection

The pump delivers 130 l/min maximum and 5.7 bar maximum pressure, with flow falling sharply above 5.0 bar.

Needed to check whether the external circuit pressure drop can be overcome at the required flow.

Source
“max. delivery 130 l/min max. delivery pressure 5.7 bar”
Selection

The expansion tank volume is 48 l and minimum filling capacity is 18 l.

Determines the minimum heat-transfer fluid charge and expansion allowance for the loop.

Source
“min. filling capacity 18 l Volume of expansion 48 l”
Safety

The refrigerant is R-449A with GHS classification H280 and GWP 1397, quantity 1.03 kg.

Affects refrigerant handling, leak response and environmental reporting obligations.

Source
“R-449A (A1, H280)”
Cumplimiento

The unit is classified as pressure equipment category Art. 4.3 PED.

Defines the pressure-equipment compliance route and documentation required at site.

Source
“Pressure equipment category Art. 4.3 PED”
Cumplimiento

The chiller is EMC Class A Group 1, which is not intended for residential environments.

Class A equipment may cause radio interference in domestic areas and needs industrial installation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The enclosure is IP20, so it must be protected from water and dust ingress at the installation site.

Prevents placing the unit in wash-down or dusty areas without additional protection.

Source
“Degree of Protection IP20”
Operation

The sound pressure level is 62 dB(A) ±4 dB(A), which matters for occupied plant rooms.

Used for acoustic planning and personnel exposure assessment.

Source
“sound pressure level +/- 4 dB(A) 62 dB(A)”
Maintenance

The unit has a digital level indicator and an atmospheric open expansion tank that require periodic checking.

Level monitoring is part of routine operation to avoid pump cavitation and air ingress.

Source
“With digital level indicator.”