Installation
The chiller requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device rating before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The unit is water-cooled and needs a cooling water connection of G1/2 male with a minimum differential pressure of 0.4 bar.
Cooling water supply pressure and connection size must be verified during site preparation.
Source
“min. cooling water differential pressure 0,4 bar”
Installation
Cooling water pressure must not exceed 6 bar, so a pressure reducer may be needed on the building water line.
Prevents damage to the cooling water circuit from overpressure.
Source
“max. cooling water pressure 6 bar”
Operation
The unit is designed for cooling water intake between 5 °C and 20 °C, and lower temperatures require insulating the cooling water pipes.
Defines the acceptable cooling water supply range and condensation protection measures.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Affects sizing of the cooling water supply and expected performance at site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in opposite directions.
Ensures the site power quality is within the allowed envelope for safe operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and the need to procure a matching plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.
Informs safety measures for refrigerant handling and room ventilation.
Source
“R-449A (A1, H280)”
Cumplimiento
The unit is classified as Class A, Group 1 for electromagnetic disturbance, which has implications for the installation environment.
Class A equipment may require additional EMC measures in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
Restricts the installation location to dry, protected indoor areas.
Source
“Degree of Protection IP20”
Operation
The ambient temperature must stay between 5 °C and 40 °C for correct operation.
Defines the allowable room temperature range at the installation site.
Source
“min. ambient temperature 5 °C”
Selection
The pump delivers a maximum of 130 l/min at 5.7 bar, but flow falls sharply at higher pressures, which matters for high-pressure-drop applications.
Helps verify the pump can overcome the pressure drop of the external circuit.
Source
“max. delivery pressure 5,7 bar”
Maintenance
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and treatment to avoid corrosion and fouling of the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”