Unichiller P075

Contenido

  1. Overview p. 1
    Product introduction and key features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 10 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Net weight309 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)10,5 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Unichiller P075 — Technical data according to DIN 12876

ParameterValue
Order-No.3040.0033.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 20°C7,5 kW
at 0°C5,3 kW
at -10°C3,3 kW
at -20°C1,8 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,035 kg
CO2 equivalent1,4 t
Circulation pump:E2
max. delivery130 l/min
max. delivery pressure5,7 bar
Delivery at 2,4 bar116 l/min
Delivery at 2,5 bar115 l/min
Delivery at 3,0 bar108 l/min
Delivery at 3,5 bar100 l/min
Delivery at 4,0 bar92 l/min
Delivery at 4,5 bar80 l/min
Delivery at 5,0 bar59 l/min
Delivery at 5,5 bar20 l/min
Pump connectionG1 1/4 male
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1165 mm

** см. сноску в оригинале (не приведена в тексте страницы)

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the site must provide a suitable three-phase feed and protection.

Electrical supply and fuse rating must be confirmed before installation to avoid rework or tripping.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must supply and fit the plug.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

Determines whether the installation environment is acceptable and what additional enclosure may be needed.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 °C and 40 °C, so the room climate and ventilation must be checked at the installation site.

Operation outside this range risks poor performance or equipment damage and must be avoided during siting.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is quoted at 20 °C room temperature and drops if ambient rises, so selection must account for the actual site ambient.

Sizing on nominal capacity alone can lead to insufficient cooling in a warm plant room.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Performance

The pump delivers 130 l/min maximum but only 20 l/min at 5.5 bar, so the system pressure drop must be matched to the pump curve.

High-pressure-drop applications need verification that the pump still provides adequate flow.

Source
“Delivery at 5,5 bar 20 l/min”
Installation

The pump connection is G1 1/4 male, so mating hoses or adapters must be ordered to match this thread.

Prevents mismatch of process connections and delays during mechanical installation.

Source
“Pump connection G1 1/4 male”
Installation

The unit has a minimum filling capacity of 18 l and an expansion volume of 48 l, so the external loop volume and expansion allowance must be planned.

Ensures the heat transfer circuit is sized correctly and the expansion tank can absorb volume changes.

Source
“min. filling capacity 18 l”
Safety

The refrigerant is R-449A with GHS hazard H280 and GWP 1397, so refrigerant handling and leak response procedures must be in place.

Affects safety documentation, service training and environmental compliance for the installation.

Source
“R-449A (A1, H280)”
Cumplimiento

The equipment is classified Class A, Group 1 for EMC, so it may cause radio interference in residential areas and installation location matters.

Class A equipment is intended for industrial environments and may not be suitable for residential use without mitigation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Voltage and frequency tolerances are ±5% and ±2% but not simultaneously in the unfavorable combination, so supply quality must be verified.

Poor supply quality can cause malfunction and must be checked against the allowed tolerance envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit weighs 309 kg, so lifting equipment and floor loading must be planned for transport and siting.

Affects rigging, access route and structural floor assessment during installation.

Source
“Net weight 309 kg”
Operation

The unit is for externally closed applications, so the process circuit must be closed and not open to atmosphere.

Using an open circuit would violate the intended application and risk contamination or overflow.

Source
“For externally closed applications.”
Maintenance

The expansion tank is atmospheric open and has a cover, so the cover must be kept in place and the tank level monitored.

Maintains proper expansion function and prevents contamination or evaporation of heat transfer fluid.

Source
“cover expansion tank”