Unichiller P075w Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller P075w chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 16 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -10°C270 l/h
Consumption at water 15°C, flow -20°C240 l/h
min. cooling water differential pressure0,5 bar
max. cooling water pressure6 bar
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1050 mm
Net weight301 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)10 A
Fuse (3 phase)3x10A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S344729

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller P075w — Technical data according to DIN 12876

ParameterValue
Order-No.3040.0011.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C at 20°C7,5 kW
Cooling power at ambient temperature +20°C at 0°C5,3 kW
Cooling power at ambient temperature +20°C at -10°C3,3 kW
Cooling power at ambient temperature +20°C at -20°C1,8 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,95 kg
CO2 equivalent1,3 t
Circulation pumpE2
max. delivery130 l/min
max. delivery pressure5,7 bar
Delivery at 2,4 bar116 l/min
Delivery at 2,5 bar115 l/min
Delivery at 3,0 bar108 l/min
Delivery at 3,5 bar100 l/min
Delivery at 4,0 bar92 l/min
Delivery at 4,5 bar80 l/min
Delivery at 5,0 bar59 l/min
Delivery at 5,5 bar20 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C348 l/h
Consumption at water 15°C, flow 0°C330 l/h

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 3x10 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device rating before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

The unit is water-cooled and needs a cooling water supply with at least 0.5 bar differential pressure and no more than 6 bar.

Cooling water pressure must be verified on site to ensure sufficient flow and avoid overpressure damage.

Source
“min. cooling water differential pressure 0,5 bar”
Installation

Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Defines the acceptable cooling water source temperature and the need for pipe insulation to prevent condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Performance must be derated if the available cooling water is warmer than the rated 15 °C.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Prevents corrosion or damage by selecting suitable water quality and avoiding incompatible additives.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the site power quality is within the allowed envelope before commissioning.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit is rated for ambient temperatures from 5 °C to 40 °C and IP20 protection, so it must be installed in a dry indoor environment.

Defines the permissible room conditions and excludes wet or dusty locations.

Source
“max. ambient temperature 40 °C”
Selection

The refrigerant is R-449A with GWP 1397 and a charge of 0.95 kg, giving a CO2 equivalent of 1.3 t.

Relevant for refrigerant regulations, leak checks and environmental compliance.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Selection

The pump delivers up to 130 l/min and 5.7 bar, but flow falls sharply at high pressure, so the system pressure drop must be checked.

Ensures the pump can overcome the external circuit pressure drop at the required flow.

Source
“max. delivery pressure 5,7 bar”
Installation

The unit needs a minimum filling capacity of 18 l and has a 48 l expansion volume, which affects the external loop volume and expansion tank sizing.

Determines the minimum heat transfer fluid volume and expansion volume for safe operation.

Source
“min. filling capacity 18 l”
Installation

The overall dimensions are 740x1160x1050 mm and the net weight is 301 kg, so space and floor loading must be planned.

Needed for site layout, access routes and structural floor loading.

Source
“Overall dimensions WxDxH ** 740x1160x1050 mm”
Safety

The safety classification is I / NFL according to DIN 12876, which defines the permitted application and safety requirements.

Confirms the unit is suitable for the intended temperature control application under DIN 12876.

Source
“Safety classification I / NFL”
Operation

The unit is designed for externally closed applications, so the external circuit must be closed.

Prevents misapplication with open baths or unsealed external circuits.

Source
“For externally closed applications.”