Product introduction and general description of the Unistat 1005w unit
Specpp. 2-3
Technical data according to DIN 12876 including performance and dimensions
p. 2table, 30 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Refrigerant 3rd stage (ASHRAE, GHS)
R-14 (A1, H280)
Global Warming Potential (GWP)
7380
Refrigerant quantity 3rd stage
0,6 kg
CO2 equivalent 3rd stage
4,4 t
UN-number
UN 3358
Circulation pump:
max. delivery
30 l/min
max. delivery pressure
0,9 bar
Delivery at 0,2 bar
27 l/min
Delivery at 0,4 bar
23 l/min
Delivery at 0,6 bar
15 l/min
Delivery at 0,8 bar
8 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
300 l/h
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
min. filling capacity
3,6 l
Filling capacity expansion tank
8 l
Overall dimensions WxDxH **
700x804x1520 mm
Net weight
340 kg
Power supply requirement
400V 3~ 50Hz
max. current
16,5 A
Fuse
3x20A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S585552
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Technical data according to DIN 12876 — Unistat 1005w (Order-No.: 1062.0002.01)
Characteristic
Value
Operating temperature range
-120...100 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
2 kW
Cooling power with Thermooil at 100°C
1,5 kW
Cooling power with Thermooil at 50°C
1,5 kW
Cooling power with Thermooil at 20°C
1,5 kW
Cooling power with Kryothermal S at 0°C
1,5 kW
Cooling power with Kryothermal S at -20°C
1,5 kW
Cooling power with Kryothermal S at -40°C
1,5 kW
Cooling power with Kryothermal S at -60°C
1,4 kW
Cooling power with Kryothermal S at -80°C
1,4 kW
Cooling power with Kryothermal S at -100°C
1 kW
Cooling power with Kryothermal S at -120°C
0,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,25 kg
CO2 equivalent
2,7 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,785 kg
CO2 equivalent 2nd stage
11,6 t
Значения приведены дословно по тексту страницы; строки 'Cooling power with Thermooil' и 'Cooling power with Kryothermal S' в оригинале разбиты на подстроки по температурам.
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 16.5 A maximum current and 3x20 A fusing, so the site electrical infrastructure must be sized accordingly.
Determines the electrical supply, cable and breaker rating needed at the installation site.
Source
“Power supply requirement 400V 3~ 50Hz max. current 16,5 A Fuse 3x20A”
Installation
The cooling water circuit needs at least 3 bar differential pressure and must not exceed 6 bar, with G1/2 male connections.
Defines the required cooling water supply pressure and connection type for proper installation.
Source
“min. cooling water differential pressure 3 bar max. cooling water pressure 6 bar”
Installation
The unit is designed for cooling water intake between 5 °C and 20 °C, and pipes must be insulated if water temperature can fall below the condensation point.
Prevents condensation and ensures the specified cooling performance is achievable.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Affects achievable cooling capacity and cooling water supply sizing at the site.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.
Avoids surprises during commissioning because the customer must supply the mains plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento
Voltage and frequency tolerances are only valid in combination: ±5% voltage with ±2% frequency, and opposite-sign deviations are not allowed.
Ensures the supply quality at the installation site is within the permitted operating envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant is R-452A (A1, H280) with a GWP of 2141 and a CO2 equivalent of 2.7 t, which matters for refrigerant handling and environmental reporting.
Relevant for refrigerant safety classification, leak response and CO2-equivalent reporting.
Source
“R-452A (A1, H280)”
Safety
The second and third refrigerant stages use R-23 and R-14 respectively, both classified A1 with H280, adding to the total CO2 equivalent.
Needed for full refrigerant inventory, safety classification and environmental compliance.
The pump delivers 30 l/min maximum and 0.9 bar maximum pressure, with flow falling to 8 l/min at 0.8 bar, so external circuit pressure drop must be checked.
Determines whether the pump can supply the external application's flow and pressure requirements.
The minimum filling capacity is 3.6 l and the expansion tank holds 8 l, which sets the heat transfer fluid charge and expansion allowance.
Guides heat transfer fluid selection and filling volume during commissioning.
Source
“min. filling capacity 3,6 l Filling capacity expansion tank 8 l”
Installation
The unit weighs 340 kg and measures 700x804x1520 mm, so floor loading and access clearance must be planned.
Needed for transport, rigging and installation space planning.
Source
“Overall dimensions WxDxH ** 700x804x1520 mm Net weight 340 kg”
Operation
The ambient temperature must stay between 5 °C and 40 °C, and the unit is rated IP20, so it must be installed in a dry, protected location.
Defines the permissible room conditions and protection against dust and water ingress.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Cumplimiento
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which has implications for the installation environment.
Class A equipment may require EMC precautions in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection
The unit is a water-cooled refrigerated heating circulator with a stainless steel pump and cooled shaft seal, suitable for externally closed and open applications.
Confirms the application suitability and cooling medium requirement at the selection stage.
Source
“Refrigerated Heating Circulator with water-cooled cooling machine and optical level indicator.”