Product introduction and general description of the Unistat 405w unit
Specpp. 1-2
Technical data and performance specifications according to DIN 12876
p. 2table, 19 rows
p. 1table, 28 rows
Technical data according to DIN 12876
Parameter
Value
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
72 l/h
Consumption at water 15°C, flow -20°C
54 l/h
Consumption at water 25°C, flow 0°C
186 l/h
Verbrauch b. Wasser 25°C, Vorlauf -20°C
186 l/h
min. cooling water differential pressure
0,5 bar
max. cooling water pressure
6 bar
min. filling capacity
1,5 l
Filling capacity expansion tank
2,8 l
Overall dimensions WxDxH **
426x327x631 mm
Net weight
55 kg
Power supply requirement
208V 2~ 60Hz
max. current
18,5 A
Fuse
20 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S585019
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-45...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
3 kW
Cooling power with Thermooil at 250°C
1,3 kW
Cooling power with Thermooil at 200°C
1,3 kW
Cooling power with Thermooil at 100°C
1,3 kW
Cooling power with Ethanol at 0°C
1,3 kW
Cooling power with Ethanol at -20°C
0,7 kW
Cooling power with Ethanol at -40°C
0,15 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
Refrigerant quantity
0,09 kg
CO2 equivalent
0,0 t
UN-number
UN 3358
Circulation pump: max. delivery
55 l/min
Circulation pump: max. delivery pressure
0,9 bar
Pump connection
M24x1,5 male
max. permissible kin. viscosity
50 mm²/s
Order-No.: 1002.0062.01; Unistat 405w
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires cooling water supply with a minimum differential pressure of 0.5 bar and must not exceed 6 bar.
Ensures the water-cooled refrigeration circuit receives adequate flow without overpressurizing the condenser.
Source
“min. cooling water differential pressure 0,5 bar”
Installation
Cooling water inlet temperature must be kept between 5°C and 20°C for rated performance.
Defines the site cooling water requirements and warns that higher inlet temperatures reduce cooling capacity.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature can fall below the dew point, the cooling water pipes inside the unit must be insulated.
Prevents condensation and corrosion inside the unit when cold water is supplied.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation
The electrical supply must be 208 V two-phase 60 Hz with a 20 A fuse and maximum current of 18.5 A.
Determines the site power infrastructure and breaker sizing needed before installation.
Source
“Power supply requirement 208V 2~ 60Hz”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Prevents motor and compressor damage from unfavorable supply conditions at the installation site.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The pump connection is M24x1.5 male and requires mating fittings per DIN 3863/3870.
Ensures correct hose or pipe adapters are ordered for the external circuit.
Source
“Pump connection M24x1,5 male”
Operation
The expansion tank is thermally passive, so no heat transfer fluid vapor or moisture absorption occurs, and for external open baths the tank is blocked off.
Allows the thermostat to be placed below or above the application level and avoids fluid degradation.
Source
“No HTF vapour and no moisture absorption because the expansion tank is thermally passive.”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.
Guides selection of a suitable heat transfer fluid that the pump can circulate at low temperatures.
Source
“max. permissible kin. viscosity 50 mm²/s”
Performance
Cooling power drops significantly at low temperatures with ethanol, from 1.3 kW at 0°C to 0.15 kW at -40°C.
Sets realistic expectations for heat removal capacity at the low end of the operating range.
Source
“at -40°C 0,15 kW”
Performance
Cooling water consumption increases with warmer supply water, e.g. 72 l/h at 15°C versus 186 l/h at 25°C for 0°C flow.
Affects sizing of the facility cooling water supply and drain capacity.
Source
“Consumption at water 25°C, flow 0°C 186 l/h”
Safety
The refrigerant is R-1270 (A3, H220) with GWP 0 and a charge of only 0.09 kg.
A3 flammability classification requires awareness of ventilation and ignition source control at the installation site.
Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”
Safety
The unit is classified as safety class III / FL according to DIN 12876.
Indicates suitability for flammable liquids and defines the safety category for the application.
Source
“Safety classification III / FL”
Cumplimiento
The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance.
Class A equipment may require additional EMC measures in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable water quality is required.
Guides water treatment to avoid corrosion and fouling of these materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”