Unistat 515w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat 515w unit
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 17 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C504 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity4,1 l
Filling capacity expansion tank3,7 l
Overall dimensions WxDxH **460x554x1455 mm
Net weight181 kg
Power supply requirement400V 3~ 50Hz
max. current14,5 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S432364

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-50...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C7 kW
Cooling power with Thermooil at 200°C7 kW
Cooling power with Thermooil at 100°C7 kW
Cooling power with Thermooil at 20°C7 kW
Cooling power with Ethanol at 0°C5,3 kW
Cooling power with Ethanol at -20°C2,8 kW
Cooling power with Ethanol at -40°C0,9 kW
Cooling power with Ethanol at -50°C0,25 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,45 kg
CO2 equivalent3,1 t
Circulation pump:MK pump
max. delivery112 l/min
max. delivery pressure1,5 bar
Delivery at 0,5 bar87 l/min
Delivery at 0,75 bar72 l/min
Delivery at 1,0 bar54 l/min
Delivery at 1,25 bar27 l/min
Pump connectionM30x1,5 male

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply and breaker/fuse rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 1 bar minimum differential pressure and no more than 6 bar, with G1/2 male connections.

Ensures the facility cooling water loop can meet the required pressure window and connection size.

Source
“min. cooling water differential pressure 1 bar”
Operation

Cooling water inlet must be kept between 5 °C and 20 °C, and pipes must be insulated if below the condensation point.

Operating outside this range reduces cooling power or causes condensation, affecting reliability.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures with ethanol, from 5.3 kW at 0 °C to 0.25 kW at -50 °C.

Critical for sizing the unit when low-temperature process cooling is required.

Source
“at -50°C 0,25 kW”
Safety

The refrigerant R-452A has a GWP of 2141 and a CO2 equivalent of 3.1 t, which matters for environmental compliance and leak reporting.

Affects refrigerant handling regulations and environmental reporting obligations.

Source
“Global Warming Potential (GWP) 2141”
Safety

Leakage current exceeds 3.5 mA, so residual current protection must be selected with that in mind.

Prevents nuisance tripping and ensures proper RCD selection during electrical installation.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices under 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Avoids delays at commissioning because the plug is not included.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The pump delivers 112 l/min max and 54 l/min at 1.0 bar, which constrains external circuit pressure drop.

Needed to verify the external loop pressure drop is compatible with available pump head.

Source
“max. delivery 112 l/min”
Installation

Ambient temperature must stay between 5 °C and 40 °C for proper operation.

Site conditions outside this range may derate or damage the unit.

Source
“max. ambient temperature 40 °C”
Cumplimiento

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the site power quality meets the unit's input requirements.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Maintenance

Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable water quality is required.

Guides water treatment and material compatibility to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The unit weighs 181 kg and measures 460x554x1455 mm, requiring adequate floor space and handling equipment.

Needed for site planning, access routes and structural loading.

Source
“Overall dimensions WxDxH ** 460x554x1455 mm”