Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply and breaker/fuse rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 1 bar minimum differential pressure and no more than 6 bar, with G1/2 male connections.
Ensures the facility cooling water loop can meet the required pressure window and connection size.
Source
“min. cooling water differential pressure 1 bar”
Operation
Cooling water inlet must be kept between 5 °C and 20 °C, and pipes must be insulated if below the condensation point.
Operating outside this range reduces cooling power or causes condensation, affecting reliability.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly at low temperatures with ethanol, from 5.3 kW at 0 °C to 0.25 kW at -50 °C.
Critical for sizing the unit when low-temperature process cooling is required.
Source
“at -50°C 0,25 kW”
Safety
The refrigerant R-452A has a GWP of 2141 and a CO2 equivalent of 3.1 t, which matters for environmental compliance and leak reporting.
Affects refrigerant handling regulations and environmental reporting obligations.
Source
“Global Warming Potential (GWP) 2141”
Safety
Leakage current exceeds 3.5 mA, so residual current protection must be selected with that in mind.
Prevents nuisance tripping and ensures proper RCD selection during electrical installation.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices under 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.
Avoids delays at commissioning because the plug is not included.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The pump delivers 112 l/min max and 54 l/min at 1.0 bar, which constrains external circuit pressure drop.
Needed to verify the external loop pressure drop is compatible with available pump head.
Source
“max. delivery 112 l/min”
Installation
Ambient temperature must stay between 5 °C and 40 °C for proper operation.
Site conditions outside this range may derate or damage the unit.
Source
“max. ambient temperature 40 °C”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Ensures the site power quality meets the unit's input requirements.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Maintenance
Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable water quality is required.
Guides water treatment and material compatibility to avoid corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
The unit weighs 181 kg and measures 460x554x1455 mm, requiring adequate floor space and handling equipment.
Needed for site planning, access routes and structural loading.
Source
“Overall dimensions WxDxH ** 460x554x1455 mm”