Unistat 525 Technical Datasheet

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat 525
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 14 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
max. permissible kin. viscosity50 mm²/s
min. filling capacity5,1 l
Filling capacity expansion tank18 l
Overall dimensions WxDxH **1290x795x1377 mm
Net weight417 kg
sound pressure level +/- 4 dB(A)56 dB(A)
Power supply factory configured (3 Phase)400V 3~ 50Hz
max. current (3 Phase)16 A
Fuse (3 phase)3x20A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S565613

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat 525 — Technical data according to DIN 12876

ParameterValue
Order-No.1051.0010.01
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C10 kW
Cooling power with Thermooil at 200°C10 kW
Cooling power with Thermooil at 100°C10 kW
Cooling power with Thermooil at 20°C10 kW
Cooling power with Ethanol at 0°C7 kW
Cooling power with Ethanol at -20°C4,2 kW
Cooling power with Ethanol at -40°C1,5 kW
Cooling power with Ethanol at -50°C0,65 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity3,9 kg
CO2-Äquvivalent8,3 t
Circulation pump:MK pump
max. delivery79 l/min
max. delivery pressure1,5 bar
Delivery at 0,5 bar63 l/min
Delivery at 0,75 bar54 l/min
Delivery at 1,0 bar43 l/min
Delivery at 1,25 bar29 l/min
Pump connectionM30x1,5 male

Значения приведены дословно по тексту страницы; таблица транспонирована в формат «характеристика — значение».

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 16 A maximum current and 3x20 A fusing, so the site electrical infrastructure must be sized accordingly.

Determines the required three-phase feed, breaker rating and cabling before installation.

Source
“Power supply factory configured (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.

Affects electrical installation scope and procurement of the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection on site.

Impacts earth connection and residual-current device choice for safe operation.

Source
“Attention: leakage current > 3,5mA”
Installation

The unit is rated IP20, so it must be installed in a dry indoor location protected from water ingress.

Restricts installation environment and enclosure requirements.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C room temperature.

Sets the allowable plant-room temperature and warns that performance derates with rising ambient.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Defines the acceptable supply quality window for reliable operation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The pump connection is M30x1.5 male with Y-bore and ball-socket fittings per DIN 3863/3870, so matching hoses and fittings must be specified.

Ensures correct mating of process hoses and avoids leaks at the circulation connections.

Source
“Pump connection M30x1,5 male”
Operation

The pump delivers 79 l/min maximum and 1.5 bar maximum pressure, with flow falling to 29 l/min at 1.25 bar, so external circuit pressure drop must be checked.

Allows verification that the external loop resistance is compatible with the pump curve.

Source
“max. delivery 79 l/min”
Operation

Minimum filling capacity is 5.1 l and the expansion tank holds 18 l, which sets the minimum heat-transfer fluid charge and expansion allowance.

Determines fluid volume needed and expansion volume for safe thermal cycling.

Source
“min. filling capacity 5,1 l”
Operation

Maximum permissible kinematic viscosity is 50 mm²/s, so the heat-transfer fluid must be selected to stay below this limit at the lowest operating temperature.

Prevents pump overload and poor circulation when choosing fluid for low-temperature operation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Safety

The refrigerant is R-452A (A1, H280) with GWP 2141 and 3.9 kg charge, giving 8.3 t CO2 equivalent, which affects refrigerant handling and environmental reporting.

Needed for refrigerant safety classification, leak response and CO2-equivalent reporting.

Source
“R-452A (A1, H280)”
Cumplimiento

The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Informs site EMC assessment and suitability for non-residential installation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

Cooling capacity with ethanol falls from 7 kW at 0 °C to 0.65 kW at -50 °C, so low-temperature duty must be checked against the actual fluid and temperature.

Prevents oversizing assumptions at low temperature where available cooling is much lower.

Source
“at -50°C 0,65 kW”
Maintenance

The unit has an optical level indicator and an optional float switch in the sight glass for extended security, supporting fluid-level monitoring.

Guides selection of level monitoring to protect the pump and heater from low fluid.

Source
“float switch in sight glass for extended security”