Unistat 615w Technical Data

Contenido

  1. Overview p. 1
    Introduces the Unistat 615w unit and its main features
  2. Spec pp. 1-2
    Performance, temperature range, and technical data per DIN 12876
  3. p. 2 table, 18 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C580 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity6,5 l
Filling capacity expansion tank26 l
Overall dimensions WxDxH **630x704x1520 mm
Net weight367 kg
sound pressure level +/- 4 dB(A)62 dB(A)
Power supply requirement400V 3~ 50Hz
max. current25 A
Fuse3x32A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S471936

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterUnistat 615w
Order-No.1074.0001.01
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C9,5 kW
Cooling power with Thermooil at 100°C9,5 kW
Cooling power with Thermooil at 20°C9,5 kW
Cooling power with Ethanol at 0°C9,5 kW
Cooling power with Ethanol at -20°C8 kW
Cooling power with Ethanol at -40°C4,6 kW
Cooling power with Ethanol at -60°C1,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,55 kg
CO2 equivalent5,5 t
Circulation pump: max. delivery82 l/min
Circulation pump: max. delivery pressure1,5 bar
Delivery at 0,25 bar75 l/min
Delivery at 0,5 bar67 l/min
Delivery at 0,75 bar59 l/min
Delivery at 1,0 bar47 l/min
Delivery at 1,25 bar29 l/min
Pump connectionM30x1,5 male

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 25 A maximum current and 3x32 A fusing, so the electrical installation must be sized accordingly.

Determines the site power distribution, cable cross-section and breaker rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 25 A Fuse 3x32A”
Installation

Cooling water must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and the circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.

Defines the facility cooling-water supply conditions and water quality compatibility required for reliable operation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

The cooling water connection is G1/2 male and the unit consumes about 580 l/h at 15 °C water and 20 °C flow, so the water supply and return lines must handle this flow.

Needed to size cooling-water piping, valves and facility supply capacity.

Source
“Cooling water connection G1/2 male Consumption at water 15°C, flow 20°C 580 l/h”
Installation

The unit is water-cooled and uses R-452A refrigerant with a GWP of 2141 and 2.55 kg charge (5.5 t CO2 equivalent), which affects refrigerant handling and environmental compliance.

Important for refrigerant regulations, leak detection and environmental permitting at the installation site.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 2,55 kg CO2 equivalent 5,5 t”
Installation

The unit weighs 367 kg net and measures 630x704x1520 mm, so floor loading and access routes must be checked before moving it into place.

Determines rigging, floor loading and doorway/access clearance requirements.

Source
“Overall dimensions WxDxH ** 630x704x1520 mm Net weight 367 kg”
Operation

The operating temperature range is -60 to 200 °C with 0.01 K display resolution and Pt100 sensors, suitable for precise external closed and open applications.

Confirms the unit meets the required process temperature span and stability for the application.

Source
“Operating temperature range -60...200 °C Temperature stability at -10°C 0,01 K”
Performance

Cooling power is 9.5 kW with Thermooil from 20 to 200 °C but drops to 1.2 kW with ethanol at -60 °C, so low-temperature duty must be checked against the load.

Prevents oversizing expectations at the cold end and ensures the process heat load is removable at -60 °C.

Source
“Cooling power with Ethanol at -60°C 1,2 kW”
Performance

The circulation pump delivers up to 82 l/min and 1.5 bar, with 47 l/min at 1.0 bar, so external circuit pressure drop must be matched to the pump curve.

Needed to verify the pump can overcome the pressure drop of the external application loop.

Source
“max. delivery 82 l/min max. delivery pressure 1,5 bar Delivery at 1,0 bar 47 l/min”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid selection at low temperatures.

Ensures the chosen heat transfer fluid remains pumpable within the unit's limits across the operating range.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The minimum filling capacity is 6.5 l and the expansion tank holds 26 l, so the external loop volume plus unit volume must be considered when charging fluid.

Determines the minimum fluid charge and expansion volume available for the external circuit.

Source
“min. filling capacity 6,5 l Filling capacity expansion tank 26 l”
Installation

The unit is rated IP20 and must be operated between 5 and 40 °C ambient, so it must be installed in a dry, temperature-controlled indoor location.

Sets the environmental envelope and ingress protection limits for the installation room.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and a -5% voltage with +2% frequency is explicitly not allowed.

Ensures the site power quality is within the unit's permitted supply envelope before commissioning.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug and connection must be provided on site.

Affects the electrical installation scope and the need to procure a matching plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit has safety classification III/FL and adjustable overtemperature protection according to DIN 12876, relevant for flammable or sensitive fluids.

Confirms the safety category for use with flammable heat transfer fluids and the required protective settings.

Source
“Safety classification III / FL”