Unistat 630w Temperature Control System

Contenido

  1. Overview p. 1
    Introduces the Unistat 630w temperature control unit
  2. Spec pp. 1-2
    Performance data, dimensions, and operating parameters per DIN 12876
  3. p. 2 table, 19 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Net weight715 kg
Power supply requirement400V 3~ 50Hz
max. current48,7 A
Fuse3x50A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S593532

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 630w (Order-No.: 5018.0100.01)

CharacteristicValue
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 200°C22 kW
Cooling power with Thermooil at 100°C22 kW
Cooling power with Thermooil at 20°C22 kW
Cooling power with Ethanol at 0°C21 kW
Cooling power with Ethanol at -20°C20 kW
Cooling power with Ethanol at -40°C10,5 kW
Cooling power with Ethanol at -60°C2,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity5,2 kg
CO2 equivalent11,1 t
UN-numberUN 2857
Circulation pump:MK pump
max. delivery196 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min

Таблица приведена как напечатана (характеристика — значение); строки Cooling power сгруппированы по теплоносителю (Thermooil / Ethanol) с указанием температуры.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a maximum current of 48.7 A and 3x50 A fuses, so the electrical installation must be sized accordingly.

Determines the site power distribution, cable cross-section and protection devices needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 48,7 A Fuse 3x50A”
Installation

The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure, with G3/4 male connections.

Defines the required cooling water supply pressure and piping connections for correct and safe operation.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Operation

The unit is designed for cooling water intake between 5 °C and 20 °C, and cooling power drops as the cooling water temperature rises.

Cooling water temperature directly affects achievable cooling capacity and must be verified against the process load.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature can fall below the condensation point, the cooling water pipes in the unit must be insulated.

Prevents condensation and corrosion on the cooling water circuit when cold water is used.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Cumplimiento

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

The site power quality must be checked; an out-of-tolerance combination can damage the unit or void compliance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The equipment has a leakage current greater than 3.5 mA, which requires appropriate protective earthing and RCD selection.

Affects earth connection design and residual current device selection for safe operation.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

The customer must plan and supply the correct plug and connection for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Selection

The refrigerant is R-452A with GWP 2141 and a charge of 5.2 kg, giving a CO2 equivalent of 11.1 t that must be considered for leak monitoring and F-gas obligations.

Determines refrigerant handling, leak-check and reporting requirements under applicable regulations.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 5,2 kg CO2 equivalent 11,1 t”
Installation

The unit is IP20 and must be operated between 5 °C and 40 °C ambient temperature, so it needs a dry, protected indoor location.

Defines the environmental limits and protection level for siting the equipment.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit weighs 715 kg and measures 918x963x1771 mm, so floor loading and access routes must be planned.

Needed for transport, floor loading and space planning at the installation site.

Source
“Overall dimensions WxDxH ** 918x963x1771 mm Net weight 715 kg”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, which limits fluid selection at low temperatures.

Ensures the pump can circulate the chosen fluid; too-viscous fluid at low temperature can overload the pump.

Source
“max. permissible kin. viscosity 100 mm²/s”
Operation

The minimum filling capacity is 8.9 l and the expansion tank holds 19.9 l, which sets the required heat transfer fluid volume.

Determines the amount of heat transfer fluid to procure and the expansion volume available.

Source
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”
Performance

Cooling power with ethanol falls from 21 kW at 0 °C to 2.5 kW at -60 °C, so the usable capacity at the lowest temperatures is much lower.

The process heat load must be checked against the reduced cooling capacity at low temperatures.

Source
“at -60°C 2,5 kW”
Installation

The pump connection is M38x1.5 male and the mating pipework must use ball sockets to DIN 3863 and sleeve nuts to DIN 3870.

Ensures correct matching hoses and fittings are ordered for the external circuit.

Source
“Pump connection M38x1,5 male”