Induztr
Resumen técnico Visor de documentos Modelos relacionados

Unistat 680w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat 680w unit
  2. Spec pp. 1-2
    Technical performance data and operating parameters per DIN 12876
  3. p. 2 table, 8 rows
  4. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
min. filling capacity93 l
Filling capacity expansion tank230 l
Overall dimensions WxDxH **4600x2600x2250 mm
Power supply requirement400V 3~ 50Hz
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 680w (Order-No.: 1067.0003.01)

CharacteristicValue
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power96 kW
Cooling power with Thermooil at 200°C130 kW
Cooling power with Thermooil at 100°C130 kW
Cooling power with Ethanol at 0°C130 kW
Cooling power with Ethanol at -20°C80 kW
Cooling power with Ethanol at -40°C59 kW
Cooling power with Ethanol at -60°C15 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
UN-numberUN 3358
Circulation pump: MK pump max. delivery600 l/min
Circulation pump: MK pump max. delivery pressure4 bar
Pump connectionDN50 PN16 (DIN EN 1092-1)
max. permissible kin. viscosity50 mm²/s
Cooling water connectionDN65 PN16
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar

Значения приведены дословно по тексту страницы; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' объединены с соответствующими температурными точками.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase feed.

Determines electrical infrastructure and whether a plug or hard wiring is needed.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so the installer must provide the final connection.

Affects planning of the electrical connection and required site hardware.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Cooling water must be supplied at 3–6 bar differential pressure, so the site water system must meet this range.

Insufficient or excessive cooling water pressure can prevent proper operation or damage the unit.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water inlet temperature must be between 5 °C and 20 °C, and pipes must be insulated if below the dew point.

Defines the permissible cooling water supply conditions and condensation protection measures.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 3 bar differential pressure, so actual capacity depends on site conditions.

Allows correct interpretation of published cooling capacity and prediction of derating on site.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 3 bar differential pressure between cooling water inlet and -outlet.”
Performance

Higher cooling water temperature reduces cooling power and may require higher cooling water flow, so capacity must be reassessed.

Prevents oversizing assumptions and ensures the cooling water circuit can handle increased flow.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides selection of cooling water quality and avoids corrosion or material damage.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the site power quality is within the unit's permissible operating envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The refrigerant is R-1270 (A3, H220) with GWP 0 and UN number UN 3358, so refrigerant safety and transport classifications apply.

Affects safety classification, handling requirements and transport documentation.

Source
“R-1270 (A3, H220)”
Installation

The pump connection is DN50 PN16 and the cooling water connection is DN65 PN16, so matching flanges and pipework are required.

Determines the size and standard of mating flanges and hoses for installation.

Source
“Pump connection DN50 PN16 (DIN EN 1092-1)”
Installation

Pump connections use bore shape Y (60°) per DIN 3863 and ball socket/sleeve nut per DIN 3863/3870, so hoses must match this standard.

Ensures correct fitting of pipework and flexible tempering hoses without leaks.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

Ambient temperature must be between 5 °C and 40 °C, so the installation location must be controlled within this range.

Out-of-range ambient temperature can impair performance or cause faults.

Source
“min. ambient temperature 5 °C”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor environment protected from water ingress.

Prevents installation in wet or dusty locations that could damage the equipment.

Source
“Degree of Protection IP20”
Maintenance

The unit has an electronic level indicator and automatic switch-over/capacity adaptation, so level and function should be monitored during operation.

Supports routine checks and early detection of fluid level or switching issues.

Source
“Refrigerated Heating Circulator with water-cooled cooling machine and electronical level indicator.”