Product introduction and general description of the Unistat 930w
Specp. 2
Technical data according to DIN 12876 standard
p. 2table, 21 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 1,0 bar
126 l/min
Delivery at 1,5 bar
100 l/min
Delivery at 2,0 bar
66 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
1300 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
12 l
Filling capacity expansion tank
29 l
Overall dimensions WxDxH **
950x1205x1650 mm
Net weight
984 kg
Power supply requirement
460V 3~ 60Hz
max. current
66 A
Fuse
3x80A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.1/25
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-90...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
32 kW
Cooling power with Thermooil at 200°C
19 kW
Cooling power with Thermooil at 100°C
19 kW
Cooling power with Thermooil at 20°C
19 kW
Cooling power with Ethanol at 0°C
20 kW
Cooling power with Ethanol at -20°C
20 kW
Cooling power with Ethanol at -40°C
20 kW
Cooling power with Ethanol at -60°C
15 kW
Cooling power with Ethanol at -80°C
5 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
5 kg
CO2 equivalent
10,7 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
2,2 kg
CO2 equivalent 2nd stage
32,6 t
Circulation pump: MK pump max. delivery
168 l/min
Circulation pump: MK pump max. delivery pressure
2,5 bar
Circulation pump: MK pump Delivery at 0,5 bar
148 l/min
Order-No.: 1082.0002.01; Unistat 930w
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a water-cooled refrigeration machine, so a cooling water supply with at least 1 bar differential pressure and no more than 6 bar must be provided at the G3/4 male connection.
Cooling water piping and pressure regulation must be designed before installation; insufficient differential pressure will prevent rated cooling.
Source
“Cooling water connection G3/4 male”
Installation
The cooling water intake temperature must be kept between 5°C and 20°C, and if it can fall below the condensation point the cooling water pipes must be insulated.
Site cooling water conditions determine whether the specified cooling power is achievable and whether pipe insulation is mandatory.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Rated output data are valid only at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, so higher water temperature reduces cooling power and may require higher flow.
Selection must account for actual site water temperature; otherwise the quoted 19–20 kW cooling capacity will not be met.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Installation
The electrical supply must be 460 V 3~ 60 Hz with a maximum current of 66 A and 3x80 A fusing, and external branch circuit protection according to UL 489 is required.
The panel and feeder must be sized for 66 A and include UL 489 branch protection; the unit is delivered without a plug for currents above 63 A.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase devices drawing more than 63 A are supplied without cable and without plug, so the customer must provide the power cable and connection.
Installation planning must include procurement and termination of the power cable, since it is not part of the delivery.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Safety
Leakage current exceeds 3.5 mA, which affects the choice of residual current protective device and earthing arrangement.
Standard 30 mA RCDs may be unsuitable; the electrical design must account for high leakage current.
Source
“Attention: leakage current > 3,5mA”
Cumplimiento
Voltage and frequency tolerances are ±5% and ±2% respectively, but a -5% voltage combined with +2% frequency is not allowed.
Supply quality at the site must be checked against this combined tolerance rule before commissioning.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit is rated IP20 and must be operated between 5°C and 40°C ambient temperature.
The installation location must be protected from water and kept within the ambient range, otherwise protection and performance are not guaranteed.
Source
“min. ambient temperature 5 °C”
Selection
The refrigerant charges are 5 kg R-452A (GWP 2141, 10.7 t CO2e) and 2.2 kg R-23 (GWP 14800, 32.6 t CO2e), which matters for F-gas reporting and leak detection obligations.
The CO2 equivalent above 5 t triggers periodic leak checks and record-keeping under F-gas regulations.
Source
“CO2 equivalent 2nd stage 32,6 t”
Installation
The pump connection is M38x1.5 male with bore shape Y (60°) to DIN 3863, and hoses must use ball socket to DIN 3863 and sleeve nut to DIN 3870.
External piping and flexible hoses must be ordered with the correct DIN fittings to avoid leaks and mismatch.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid selection at low temperatures.
The chosen thermal fluid must stay below this viscosity over the operating range or pump delivery and cooling will suffer.
Source
“max. permissible kin. viscosity 50 mm²/s”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.
Water quality must be controlled to prevent corrosion and fouling of these materials in the cooling circuit.
Source
“Please use suitable cooling water.”
Selection
The unit has a minimum filling capacity of 12 l and a 29 l expansion tank, which sets the minimum fluid charge and expansion allowance.
The external loop volume and fluid expansion must be checked against these values during system design.
Source
“min. filling capacity 12 l”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
The overtemperature cut-out must be set for the application and is a key safety requirement for unattended operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”