Unistat GL 540w Temperature Control System Datasheet
Contenido
Overviewp. 1
Product introduction and general description of the Unistat GL 540w unit
Specpp. 1-2
Technical data, performance figures and operating parameters per DIN 12876
p. 2table, 13 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
918x963x1771 mm
Power supply requirement
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
-50...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 200°C
28 kW
Cooling power with Thermooil at 100°C
28 kW
Cooling power with Thermooil at 20°C
28 kW
Cooling power with Ethanol at 0°C
25 kW
Cooling power with Ethanol at -20°C
17 kW
Cooling power with Ethanol at -40°C
6,8 kW
Cooling power with Ethanol at -50°C
4 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)
Global Warming Potential (GWP)
1
Circulation pump: MK pump max. delivery
196 l/min
Circulation pump: MK pump max. delivery pressure
2,5 bar
Delivery at 0,5 bar
175 l/min
Delivery at 1,0 bar
152 l/min
Delivery at 1,5 bar
123 l/min
Delivery at 2,0 bar
86 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply, so the site must provide a 3~ 50 Hz feed sized for the 12 kW heater and 28 kW cooling load.
Electrical supply and protection must be planned before installation; single-phase power is not acceptable.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure, so the building water system must be checked against these limits.
Insufficient differential pressure reduces cooling capacity; excessive pressure can damage the water circuit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation occurs, which affects chiller/water supply design.
Cooling performance ratings are valid only within this water temperature window; outside it, capacity drops and condensation risk rises.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
The published cooling powers are valid only at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, so selection must account for actual site conditions.
If site water or ambient conditions differ, the 28 kW cooling power cannot be assumed.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and - outlet.”
Installation
The pump connection is M38x1.5 male and the cooling water connection is G3/4 male, so mating hoses and adapters must be ordered accordingly.
Mechanical interfaces determine which hoses, fittings and adapters are needed at installation.
Source
“Pump connection M38x1,5 male”
Operation
The circulation pump delivers 196 l/min at zero head but only 86 l/min at 2.0 bar, so external circuit pressure drop directly limits achievable flow.
External loop resistance must be calculated to confirm the required flow is available at the application.
Source
“max. delivery 196 l/min”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, which constrains fluid selection at low temperatures.
Using a too-viscous fluid at low temperature can overload the pump and reduce heat transfer.
Source
“max. permissible kin. viscosity 100 mm²/s”
Safety
The unit uses R-744 (CO2) refrigerant with GWP 1 and is water-cooled, which affects refrigerant handling and ventilation requirements.
R-744 is a high-pressure refrigerant; service and room ventilation must be planned for it.
Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280)”
Safety
The safety classification is III / FL and overtemperature protection is adjustable per DIN 12876, so the protection settings must be configured for the application.
Class III/FL determines the required safety strategy and protection settings for flammable or high-temperature operation.
Source
“Safety classification III / FL”
Installation
The unit is IP20 and must be installed at ambient temperatures between 5 °C and 40 °C, so it cannot be placed in a wet or condensing environment.
IP20 offers no water protection; installation location must be dry and within the ambient temperature range.
Source
“Degree of Protection IP20”
Installation
The overall dimensions are 918x963x1771 mm, so access and clearance for a unit of this footprint and height must be reserved.
Space planning and transport route must accommodate the unit's physical size.
Source
“Overall dimensions WxDxH ** 918x963x1771 mm”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are explicitly not allowed.
Site power quality must be verified; out-of-tolerance supply can void performance or damage the unit.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the final connection.
The electrical installer must plan for a suitable disconnect and plug or hard-wired connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The minimum filling capacity is 8.9 l and the expansion tank holds 19.9 l, which sets the minimum fluid charge and expansion allowance.
Fluid volume and expansion tank sizing must be checked against the external loop volume.