Product introduction and general description of the Unistat GL 560w
Specpp. 1-2
Technical specifications and performance data according to DIN 12876
p. 2table, 13 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
918x963x1771 mm
Power supply requirement
400V 3~ 50Hz
Pressure equipment category
II
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-50...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
24 kW
Cooling power with Thermooil at 200°C
41 kW
Cooling power with Thermooil at 100°C
41 kW
Cooling power with Thermooil at 20°C
41 kW
Cooling power with Ethanol at 0°C
37 kW
Cooling power with Ethanol at -20°C
30 kW
Cooling power with Ethanol at -40°C
15 kW
Cooling power with Ethanol at -50°C
8 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)
Global Warming Potential (GWP)
1
Circulation pump: MK pump max. delivery
196 l/min
Circulation pump: MK pump max. delivery pressure
2,5 bar
Delivery at 0,5 bar
175 l/min
Delivery at 1,0 bar
152 l/min
Delivery at 1,5 bar
123 l/min
Delivery at 2,0 bar
86 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase 50 Hz supply, so the electrical installation must be planned for three-phase power.
Determines the site power infrastructure and whether a three-phase connection is available.
Source
“Power supply requirement
400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar, so the building water system must be checked against these limits.
Incorrect cooling water pressure can prevent rated cooling performance or damage the unit.
Source
“min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake between 5 °C and 20 °C, and pipes must be insulated if water temperature can fall below the condensation point.
Cooling water temperature outside this range affects cooling capacity and may cause condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Rated cooling power is valid only at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, so site conditions must match these for the published data.
Prevents overestimating cooling capacity when site conditions differ from the rating basis.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility to avoid corrosion or fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
The pump connection is M38x1.5 male and the cooling water connection is G3/4 male, so matching hoses and fittings must be ordered.
Ensures correct connection hardware is available before installation.
Source
“Pump connection
M38x1,5 male”
Installation
The cooling water connection is G3/4 male, so the site must provide a matching G3/4 connection.
Avoids delays and leaks from mismatched cooling water fittings.
Source
“Cooling water connection
G3/4 male”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, so fluid selection must respect this limit.
Too-viscous fluid can overload the pump and reduce circulation performance.
Source
“max. permissible kin. viscosity
100 mm²/s”
Installation
The unit has IP20 protection and must be installed in an environment between 5 °C and 40 °C ambient temperature.
IP20 is not water-protected, so the installation location must be dry and within the ambient temperature range.
Source
“Degree of Protection
IP20”
Installation
The ambient temperature must be between 5 °C and 40 °C, which constrains the installation room conditions.
Operation outside this range may affect performance and reliability.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The unit uses refrigerant R-744 with GWP 1, which is relevant for environmental compliance and refrigerant handling.
Low-GWP refrigerant simplifies environmental compliance compared with high-GWP alternatives.
Source
“Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)”
Installation
The unit is water-cooled with a natural refrigerant, so a cooling water supply is mandatory for operation.
Confirms that no air-cooled option exists and cooling water must be available.