Unistat P430w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat P430w unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 21 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,5 bar63 l/min
Delivery at 2,0 bar48 l/min
Delivery at 2,5 bar30 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity4 l
Filling capacity expansion tank3,7 l
Overall dimensions WxDxH **460x570x1628 mm
Net weight182 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply requirement400V 3~ 50Hz
max. current14,7 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S603596

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterUnistat P430w
Order-No.1069.0034.01
Operating temperature range-40...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power4 kW
Cooling power with Thermooil at 250°C3,5 kW
Cooling power with Thermooil at 200°C3,5 kW
Cooling power with Thermooil at 100°C3,5 kW
Cooling power with Thermooil at 20°C3,5 kW
Cooling power with Ethanol at 0°C3,5 kW
Cooling power with Ethanol at -20°C2 kW
Cooling power with Ethanol at -40°C0,15 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,4 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump:MK pump
max. delivery97 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar86 l/min
Delivery at 1,0 bar75 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply and protection device rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit needs at least 1 bar differential pressure and must not exceed 6 bar, so the building water supply must be checked.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Out-of-range cooling water temperature reduces cooling power and risks condensation damage.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops sharply at low temperatures with ethanol, from 3.5 kW at 0 °C to only 0.15 kW at -40 °C.

Applications near the lower temperature limit must account for very low available cooling capacity.

Source
“at -40°C 0,15 kW”
Selection

The pump delivers 97 l/min at zero head but only 30 l/min at 2.5 bar, so high-pressure-drop loops need careful pump curve matching.

Ensures the selected pump can meet the flow demand of the external circuit at its actual pressure drop.

Source
“max. delivery 97 l/min”
Safety

The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, which affects site safety classification and ventilation requirements.

Flammable refrigerant requires appropriate room ventilation and ignition-source control during installation and operation.

Source
“R-1270 (A3, H220)”
Safety

Leakage current exceeds 3.5 mA, so protective earthing and residual current device selection must be verified.

High leakage current affects RCD type/rating and earthing design for safe operation.

Source
“Attention: leakage current > 3,5mA”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not combine in the unfavorable direction.

Site power quality must be checked to avoid operating outside the allowed supply envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

Prevents delays at commissioning because the customer must supply the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a wet or hot environment.

Defines the permissible installation location and protection against dust and water ingress.

Source
“Degree of Protection IP20”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or foul the cooling circuit and reduce performance.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid choice at low temperatures.

Selecting a fluid that is too viscous at low temperature will overload the pump and reduce flow.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863, so matching hoses and fittings must be ordered.

Ensures correct mating fittings and avoids leaks or delays during plumbing.

Source
“Pump connection M30x1,5 male”
Cumplimiento

The equipment is EMC class A group 1, which is intended for industrial environments and may cause interference in residential areas.

Affects where the unit may be installed and what EMC measures may be needed.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”