Product introduction and general description of the Unistat P530w unit
Specp. 2
Technical data according to DIN 12876 including performance parameters
p. 2table, 25 rows
p. 1table, 29 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 2,0 bar
154 l/min
Delivery at 3,0 bar
124 l/min
Delivery at 4,0 bar
87 l/min
Delivery at 5,0 bar
42 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
830 l/h
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,2 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
730x804x1738 mm
Net weight
439 kg
sound pressure level +/- 4 dB(A)
62 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
31,8 A
Fuse
3x32A
Pressure equipment category
II
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S608826
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unistat P530w (Order-No.: 5002.0101.01)
Characteristic
Value
Operating temperature range
-55...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 250°C
21 kW
Cooling power with Thermooil at 200°C
21 kW
Cooling power with Thermooil at 100°C
21 kW
Cooling power with Thermooil at 20°C
21 kW
Cooling power with Ethanol at 0°C
16 kW
Cooling power with Ethanol at -20°C
9 kW
Cooling power with Ethanol at -40°C
3 kW
Cooling power with Ethanol at -50°C
1 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
Refrigerant quantity
1,35 kg
CO2 equivalent
0,0 t
Circulation pump: MK pump max. delivery
201 l/min
Circulation pump: MK pump max. delivery pressure
5,3 bar
Circulation pump: MK pump Delivery at 1,0 bar
179 l/min
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 32 A fuse per phase, so the electrical installation must be sized accordingly.
Determines the site power supply, cable cross-section and protection device needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar, with 830 l/h consumption at 15 °C.
Sizing of the cooling water supply, pump and piping must meet these pressure and flow limits.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water intake must stay between 5 °C and 20 °C, and pipes must be insulated if water can drop below the condensation point.
Out-of-range cooling water reduces cooling power and can cause condensation and corrosion.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops sharply with ethanol at low temperatures (16 kW at 0 °C down to 1 kW at -50 °C), so low-temperature duty must be checked against the curve.
Selection for sub-zero processes must account for the reduced cooling power at the required temperature.
Source
“at -50°C 1 kW”
Safety
The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, so the installation area must respect flammability requirements.
A3 classification affects room ventilation, ignition-source control and site safety assessment.
Source
“R-1270 (A3, H220)”
Safety
Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must be verified.
High leakage current influences the choice of residual current device and earthing arrangement.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.
The customer must procure and fit the mains plug, affecting installation planning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the same direction, e.g. -5 % voltage with +2 % frequency is not allowed.
Supply quality at the site must be checked to avoid operating outside the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
The pump delivers 201 l/min at zero head but only 42 l/min at 5 bar, so high-pressure-drop loops must be checked against the pump curve.
Ensures the circulation pump can meet the flow demand of the external circuit at its pressure drop.
Source
“max. delivery pressure 5,3 bar”
Installation
Pump connections are M38x1.5 male with bore shape Y (60°) per DIN 3863, so matching hoses and fittings must be ordered.
Determines the hose/fitting specification needed for connecting the external circuit.
Source
“Pump connection M38x1,5 male”
Operation
Minimum filling capacity is 8.2 l and the expansion tank holds 19.9 l, which sets the heat-transfer fluid charge and expansion allowance.
Needed to calculate fluid volume, expansion volume and cost of heat-transfer fluid.
Source
“min. filling capacity 8,2 l”
Installation
The unit is IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a wet or hot environment.
Site conditions and enclosure placement must respect the IP rating and ambient limits.
Source
“Degree of Protection IP20”
Cumplimiento
The equipment is EMC class A group 1, so it is intended for industrial environments and may cause interference in residential areas.
Affects site selection and EMC coordination with sensitive equipment nearby.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Water quality must be controlled to avoid corrosion and fouling of these materials.