Unistat P540w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat P540w unit
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 24 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1480 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity9,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight473 kg
Power supply requirement400V 3~ 50Hz
max. current56 A
Fuse3x63A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S553198

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 250°C30 kW
Cooling power with Thermooil at 200°C30 kW
Cooling power with Thermooil at 100°C30 kW
Cooling power with Thermooil at 20°C30 kW
Cooling power with Ethanol at 0°C30 kW
Cooling power with Ethanol at -20°C16 kW
Cooling power with Ethanol at -40°C4 kW
Cooling power with Ethanol at -50°C1,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity3,6 kg
CO2 equivalent7,7 t
Circulation pump:MK pump
max. delivery201 l/min
max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V 3~ 50 Hz supply with a maximum current of 56 A and must be protected by 3x63 A fuses.

Electrical supply and protection must be sized accordingly before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 56 A Fuse 3x63A”
Installation

The cooling water circuit must be supplied at 5–20 °C with a differential pressure between 1 and 4 bar and a maximum pressure of 6 bar.

Cooling water supply pressure and temperature must be verified to avoid performance loss or damage.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Installation

If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.

Prevents condensation and corrosion on cold cooling water piping.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.

Water quality must be compatible with these materials to avoid corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Installation

Three-phase devices with current consumption less than 63 A are delivered with cable but without plug.

The installer must provide a suitable plug or hard-wired connection on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The equipment has a leakage current greater than 3.5 mA, which affects protective earthing and RCD selection.

Earthing and residual current protection must be designed for high leakage current.

Source
“Attention: leakage current > 3,5mA”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Supply quality must be checked to stay within the permitted voltage/frequency envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The unit is designed for cooling water intake temperatures between 5 °C and 20 °C, and higher water temperatures reduce cooling power.

Cooling water temperature directly affects achievable cooling capacity.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Selection

The pump delivers 201 l/min maximum and 5.3 bar maximum pressure, with delivery dropping to 42 l/min at 5.0 bar.

Pump curve must be matched to the pressure drop of the external circuit.

Source
“max. delivery 201 l/min max. delivery pressure 5,3 bar”
Selection

Cooling power with ethanol falls from 30 kW at 0 °C to 1.5 kW at -50 °C, so low-temperature capacity is limited.

Low-temperature applications must account for strongly reduced cooling capacity.

Source
“at -50°C 1,5 kW”
Selection

The refrigerant is R-452A with GWP 2141 and a charge of 3.6 kg, giving 7.7 t CO2 equivalent.

Refrigerant charge and CO2 equivalent matter for environmental compliance and service.

Source
“Refrigerant quantity 3,6 kg CO2 equivalent 7,7 t”
Installation

The pump connection is M38x1.5 male and pipework must use ball socket per DIN 3863 and sleeve nut per DIN 3870.

Correct mating fittings must be procured for leak-free connection.

Source
“Pump connection M38x1,5 male”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.

Fluid selection must respect viscosity limit to avoid pump overload.

Source
“max. permissible kin. viscosity 100 mm²/s”
Installation

The unit weighs 473 kg and measures 730x804x1738 mm, requiring adequate floor space and handling equipment.

Site access, floor loading and lifting provisions must be planned.

Source
“Overall dimensions WxDxH ** 730x804x1738 mm Net weight 473 kg”