Unistat P815 Temperature Control System Datasheet

Contenido

  1. Overview p. 1
    Product introduction and general description of Unistat P815
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 22 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
max. delivery67 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar61 l/min
Delivery at 1,0 bar55 l/min
Delivery at 1,5 bar48 l/min
Delivery at 2,0 bar39 l/min
Delivery at 2,5 bar28 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity3,8 l
Filling capacity expansion tank3,3 l
Overall dimensions WxDxH **460x604x1465 mm
Net weight241 kg
sound pressure level +/- 4 dB(A)64 dB(A)
Power supply factory configured (3 Phase)400V 3~ 50Hz
max. current (3 Phase)12 A
Fuse (3 phase)3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S290580

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat P815 — Technical data according to DIN 12876

ParameterValue
Order-No.1053.0009.01
Operating temperature range-85...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power2 kW
Cooling power with Thermooil at 250°C1,3 kW
Cooling power with Thermooil at 200°C1,3 kW
Cooling power with Thermooil at 100°C1,3 kW
Cooling power with Ethanol at 0°C1,5 kW
Cooling power with Ethanol at -20°C1,5 kW
Cooling power with Ethanol at -40°C1,4 kW
Cooling power with Ethanol at -60°C1,2 kW
Cooling power with Ethanol at -80°C0,2 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,8 kg
CO2-Äquvivalent3,9 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage0,35 kg
CO2-Äquvivalent 2. Stufe5,2 t
Circulation pump:MK pump

Значения приведены дословно по тексту страницы; строки Cooling power сгруппированы по теплоносителю (Thermooil / Ethanol) и температуре.

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a refrigerated heating circulator suitable for both externally closed and externally open applications, which determines whether an isolation sleeve may be needed.

Open external applications require additional accessories such as an isolation sleeve, affecting selection and cost.

Source
“As well as for externally closed and also externally open applications.”
Selection

The 'P' model pump is designed for high pressure drops, making it suitable for flow-through chemistry or semiconductor applications.

Confirms the pump variant matches high-pressure-drop loops; a standard pump may be insufficient.

Source
“Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation

The unit requires a 3-phase 400 V supply with 12 A max current and 3x16 A fusing, so the electrical installation must be sized accordingly.

Determines breaker, cable and supply requirements before installation.

Source
“max. current (3 Phase) 12 A”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the plug.

Affects site preparation and commissioning; a plug must be sourced and installed.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863, requiring matching ball-socket hoses and sleeve nuts per DIN 3870.

Ensures correct hose/adaptor selection and leak-free connection at commissioning.

Source
“Pump connection M30x1,5 male”
Operation

Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so performance must be derated in warm rooms.

Prevents undersizing cooling capacity in installations with ambient above 20 °C.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The permissible ambient temperature range is 5 to 40 °C, limiting where the unit can be installed and operated.

Site conditions outside this range void performance assumptions and may affect reliability.

Source
“max. ambient temperature 40 °C”
Safety

Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.

Affects electrical safety design; standard 30 mA RCDs may nuisance-trip or be unsuitable.

Source
“Attention: leakage current > 3,5mA”
Safety

The refrigerant R-452A is classified A1 with GHS H280 (pressurised gas), and the second stage R-23 is also A1/H280, so handling and ventilation precautions apply.

Informs refrigerant safety classification and site ventilation/emergency planning.

Source
“R-452A (A1, H280)”
Cumplimiento

The equipment is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Determines EMC environment suitability and possible need for filtering or separation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Cumplimiento

Voltage and frequency tolerances are ±5% and ±2% respectively, but certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.

Prevents operation on marginal supplies that could damage the compressor or electronics.

Source
“-5% voltage and +2% frequency -> not allowed!”
Maintenance

The refrigeration machine is air-cooled and CFC/HCFC-free, so adequate air flow and condenser cleaning are required.

Air-cooled condensers need clearance and periodic cleaning; poor airflow reduces capacity.

Source
“air-cooled, CFC- and HCFC-free”
Performance

Cooling capacity drops sharply at low temperature with ethanol, from 1.2 kW at -60 °C to 0.2 kW at -80 °C, so low-temperature loads must be small.

Critical for sizing low-temperature processes near the -85 °C limit.

Source
“at -80°C 0,2 kW”
Selection

The pump delivers 67 l/min max and 3 bar max pressure, with flow falling to 28 l/min at 2.5 bar, so system pressure drop must be checked against the pump curve.

Ensures the circulation loop can achieve required flow at the actual pressure drop.

Source
“max. delivery pressure 3 bar”