Unistat P912w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the unit
  2. Spec pp. 2-3
    Technical data according to DIN 12876 standard
  3. p. 2 table, 27 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Circulation pumpMK pump
max. delivery110 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar89 l/min
Delivery at 1,0 bar75 l/min
Delivery at 1,5 bar60 l/min
Delivery at 2,0 bar43 l/min
Delivery at 2,5 bar24 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C516 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,9 l
Filling capacity expansion tank6,5 l
Overall dimensions WxDxH **630x704x1565 mm
Net weight325 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply requirement460V 3~ 60Hz
max. current20 A
Fuse3x25A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S607523

** Overall dimensions WxDxH

Technical data according to DIN 12876

CharacteristicUnistat P912w
Order-No.1055.0013.01
Operating temperature range-90...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C7 kW
Cooling power with Thermooil at 200°C7 kW
Cooling power with Thermooil at 100°C7 kW
Cooling power with Ethanol at 0°C7 kW
Cooling power with Ethanol at -20°C7 kW
Cooling power with Ethanol at -40°C6 kW
Cooling power with Ethanol at -60°C3,5 kW
Cooling power with Ethanol at -80°C0,9 kW
Cooling power with Ethanol at -90°C0,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2 kg
CO2 equivalent4,3 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage1,1 kg
CO2 equivalent 2nd stage16,3 t

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 460V 3-phase 60Hz supply with a 20 A maximum current and 3x25A fusing, so the site electrical infrastructure must be sized accordingly.

Determines the electrical supply, breaker and cable sizing needed before installation.

Source
“Power supply requirement 460V 3~ 60Hz max. current 20 A Fuse 3x25A”
Installation

The cooling water circuit needs a minimum differential pressure of 1 bar and must not exceed 6 bar, and the unit consumes 516 l/h at 15°C water and 0°C flow.

Sizes the cooling water supply, pump and piping; insufficient differential pressure will prevent rated performance.

Source
“Consumption at water 15°C, flow 0°C 516 l/h min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The unit is designed for cooling water intake between 5°C and 20°C, and cooling water pipes must be insulated if the water temperature can fall below the condensation point.

Defines acceptable cooling water conditions and condensation protection required at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase, so performance ratings are only valid at the stated reference conditions.

Prevents overestimating available cooling capacity when site cooling water is warmer than 15°C.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion or fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.

Defines the acceptable supply quality window for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% Example -5% voltage and +2% frequency -> not allowed!”
Safety

The unit has a leakage current greater than 3.5 mA and requires external branch circuit protection according to UL 489.

Mandates additional external branch protection and affects earth leakage protection design.

Source
“Attention: leakage current > 3,5mA External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.

Avoids delays at commissioning because the customer must supply and fit the plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The pump connection is M30x1.5 male with a bore shape Y (60°) per DIN 3863, and pipework or flexible hoses must use a ball socket per DIN 3863 and sleeve nut per DIN 3870.

Specifies the exact mating fittings needed for leak-free connection of the circulation loop.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid selection at low temperatures.

Constrains the choice of heat transfer fluid, especially for low-temperature operation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Selection

The unit is water-cooled with CFC- and HCFC-free refrigerants R-452A and R-23, with a combined CO2 equivalent of about 20.6 t, which matters for refrigerant reporting and room ventilation requirements.

Affects environmental compliance, refrigerant leak reporting and site ventilation planning.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 2 kg CO2 equivalent 4,3 t”
Operation

The unit must be operated within an ambient temperature range of 5°C to 40°C and has a sound pressure level of 65 dB(A).

Sets the allowable room conditions and acoustic environment for installation.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C sound pressure level +/- 4 dB(A) 65 dB(A)”
Selection

The unit has a minimum filling capacity of 3.9 l and an expansion tank filling capacity of 6.5 l, which must be considered when sizing the external circuit.

Determines the minimum heat transfer fluid volume and expansion volume for the loop.

Source
“min. filling capacity 3,9 l Filling capacity expansion tank 6,5 l”
Installation

The unit has degree of protection IP20 and overall dimensions of 630x704x1565 mm with a net weight of 325 kg, so space and floor loading must be planned.

Needed for room layout, access and floor load planning.

Source
“Overall dimensions WxDxH ** 630x704x1565 mm Net weight 325 kg”