Unistat T320 HT Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat T320 HT
  2. Spec pp. 1-2
    Technical data according to DIN 12876 including performance parameters
  3. p. 2 table, 2 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
max. ambient temperature40 °C
from Serial-No.:1.0/25

Unistat T320 HT — Technical data according to DIN 12876

ParameterValue
Order-No.1083.0023.01
Operating temperature range65... 300 °C
Temperature stability at 70°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power at 300°C10 kW
Cooling power at 200°C10 kW
Cooling power at 100°C3,5 kW
Circulation pump:MK pump
max. delivery96 l/min
max. delivery pressure3,5 bar
Delivery at 0,5 bar82 l/min
Delivery at 1,0 bar72 l/min
Delivery at 1,5 bar60 l/min
Delivery at 2,0 bar45 l/min
Delivery at 2,5 bar33 l/min
Delivery at 3,0 bar20 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity7 l
Filling capacity expansion tank24 l
Overall dimensions WxDxH **540x749x1700 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C

** The dimensions specified below are approximate and may be subject to change.

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a heating circulator without active refrigeration, so 65 °C minimum can only be reached with external heat loss or water cooling.

Prevents specifying this unit for cooling duties below 65 °C without a suitable heat sink or water cooling.

Source
“These devices do not have an active refrigeration system.”
Installation

Three-phase devices below 63 A are supplied with cable but no plug, so the installer must provide a suitable plug.

Affects electrical installation planning and bill of materials for the power connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are coupled: ±5 % voltage only with ±2 % frequency, and certain combinations are not allowed.

Needed to verify site power quality before commissioning the 400 V three-phase unit.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The pump connection is M30x1.5 male with DIN 3863 ball-socket and DIN 3870 sleeve-nut mating, so hoses must match this standard.

Ensures correct selection of tempering hoses and fittings for leak-free connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation

The unit requires 540x749x1700 mm space and the manufacturer explicitly requires respecting space requirements.

Needed for room layout, access, and ventilation planning during installation.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Operation

Ambient temperature must stay between 5 °C and 40 °C for proper operation.

Defines the allowable site environment and prevents operation outside the specified range.

Source
“min. ambient temperature 5 °C”
Operation

For air-cooled operation the application heat loss must be between 0.1 kW and 2.8 kW depending on the unit.

Determines whether the process load is compatible with air-cooled operation.

Source
“Air cooled: The heat loss must be between 0.1kW and 2.8kW, depending on the unit.”
Operation

To reach 65 °C in air-cooled mode the ambient temperature must be well below 60 °C.

Affects achievable minimum temperature and site ambient conditions.

Source
“In order to reach 65°C the ambient temperature must be well below 60°C.”
Operation

The maximum permissible kinematic viscosity of the fluid is 50 mm²/s.

Guides selection of thermofluid to avoid pump overload and poor circulation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Operation

Minimum filling capacity is 7 l and the expansion tank holds 24 l.

Needed to size the fluid charge and expansion volume for the loop.

Source
“min. filling capacity 7 l”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Confirms safety compliance and the need to set the overtemperature limit during commissioning.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Cumplimiento

The unit is classified Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require EMC precautions in residential or light-industrial environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Performance

Cooling power drops from 10 kW at 200–300 °C to 3.5 kW at 100 °C, which matters for load sizing.

Prevents overestimating available cooling capacity at lower operating temperatures.

Source
“Cooling power at 100°C 3,5 kW”
Maintenance

The expansion tank is not thermoregulated and can be locked for open baths.

Affects fluid handling and configuration depending on closed-system or open-bath use.

Source
“Expansion tank (not thermoregulated) for closed systems, lockable for open baths.”