  1. Overview p. 1
    Product introduction and general description of Unistat tango
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 12 rows
  4. p. 1 table, 28 rows

Technical data according to DIN 12876

ParameterValue
min. filling capacity1,5 l
Filling capacity expansion tank2,8 l
Overall dimensions WxDxH **426x327x631 mm
Net weight57 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current15,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S585146

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-45...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power3 kW
Cooling power with Thermooil at 250°C0,7 kW
Cooling power with Thermooil at 200°C0,7 kW
Cooling power with Thermooil at 100°C0,7 kW
Cooling power with Ethanol at 0°C0,7 kW
Cooling power with Ethanol at -20°C0,4 kW
Cooling power with Ethanol at -40°C0,06 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,15 kg
CO2-Äquvivalent0,0 t
UN-numberUN 3358
Circulation pump: max. delivery55 l/min
Circulation pump: max. delivery pressure0,9 bar
Pump connectionM24x1,5 male
max. permissible kin. viscosity50 mm²/s

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit covers a very wide operating temperature range from -45 to 250 °C, so it can serve both deep-cold and high-temperature external circuits with one circulator.

Defines whether the thermostat can cover the full process window without a second unit.

Source
“Operating temperature range -45...250 °C”
Safety

The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, so the installation location and ignition-source assessment must account for an A3 refrigerant charge.

A3 flammability affects room classification, ventilation and permitting during selection and installation.

Source
“R-1270 (A3, H220)”
Installation

The refrigerant charge is only 0.15 kg, which keeps the flammable charge small but still requires the UN 3358 transport and handling rules to be respected.

Charge size drives refrigerant safety measures and transport documentation.

Source
“Refrigerant quantity 0,15 kg”
Performance

Cooling capacity collapses to 0.06 kW at -40 °C with ethanol, so the heat load at the cold end must be kept very low or the setpoint will not be held.

Prevents oversizing the process load at the low-temperature end of the range.

Source
“at -40°C 0,06 kW”
Installation

The pump connection is M24x1.5 male, so hoses or adapters must match this thread; an M16x1 male to M24x1.5 female adaptor is supplied.

Determines which hoses and fittings are needed for the external circuit.

Source
“Pump connection M24x1,5 male”
Operation

Maximum permissible kinematic viscosity is 50 mm²/s, so the heat transfer fluid must be chosen to stay below this limit across the whole operating temperature range.

Guides HTF selection and avoids pump overload or poor circulation at low temperature.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The expansion tank is thermally passive and is blocked off for external open baths, allowing the thermostat to be mounted below or above the application level.

Gives freedom in physical placement of the unit relative to the open bath.

Source
“For external open baths the expansion tank will be blocked off.”
Installation

Minimum filling capacity is 1.5 l and the expansion tank holds 2.8 l, so the external loop volume plus the unit must be planned within these limits.

Sets the minimum HTF charge and helps size the expansion volume for the loop.

Source
“min. filling capacity 1,5 l”
Installation

The unit needs a 220-240 V single- or two-phase supply with a maximum current of 15.5 A and a 16 A fuse, so the site power feed must be sized accordingly.

Electrical supply and protection must be confirmed before installation.

Source
“max. current 15,5 A”
Installation

The enclosure is IP20, so the unit must be installed in a dry indoor location protected from water ingress.

Rules out washdown or outdoor locations and affects room selection.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 and 40 °C, and above 20 °C the cooling capacity may drop, so the plant room must be kept cool.

Explains derating and the need for adequate room ventilation.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and a -5% voltage with +2% frequency combination is explicitly not allowed.

Prevents connecting the unit to a supply whose combined voltage/frequency deviation exceeds the EN60034-1 window.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the site plug.

Affects electrical installation scope and ordering of the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento

The equipment is EMC class A group 1 to EN55011, so it is intended for industrial environments and may cause interference in residential areas.

Determines EMC environment suitability and possible filtering needs.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”