Induztr
Resumen técnico Visor de documentos Modelos relacionados

IDP-15 Dry Scroll Vacuum Pump Instructions for Use

Contenido

  1. Quality p. 3
    Manufacturer warranty terms and conditions
  2. Overview p. 5
    Table of contents for the manual
  3. Overview pp. 7-11
    General safety and usage instructions
  4. Procedure pp. 12-15
    Pump installation and connection procedures
  5. Procedure pp. 16-22
    Operating procedures and controls
  6. Procedure pp. 23-30
    Routine maintenance and service procedures
  7. Procedure pp. 31-38
    Fault diagnosis and corrective actions
  8. Spec pp. 39-48
    Technical specifications and performance data
  9. Ordering pp. 49-50
    Spare parts list and ordering information
  10. p. 54
    Contact details for service and support
  11. p. 12 table, 22 rows
  12. p. 13 table, 8 rows
  13. p. 23 table, 9 rows
  14. p. 49 table, 3 rows
  15. p. 25 table, 2 rows
  16. p. 29 table, 9 rows
  17. p. 31 table, 2 rows
  18. p. 39 table, 13 rows
  19. p. 40 table, 3 rows
  20. p. 50 table, 40 rows

Table 1 Specifications

SpecificationIDP-15 Dry Scroll Single Hermetic Vacuum Pump
ModelIDP-15 Dry Scroll Single Hermetic Vacuum Pump
Interface dimensionsSee Figure 1 on page 14
Peak pumping speed❑ 60 Hz: 257 L/m, 15.4 m3/hr ❑ 50 Hz: 213 L/m, 12.8 m3/hr
MediaNo corrosive, explosive or particulate forming gases
Ultimate pressure1 x 102 Torr (1.3 x 102 mbar)
Maximum inlet pressure1.0 atmosphere (0 psig)
Maximum outlet pressure6.5 psig
Maximum recommended continuous inlet pressure300 Torr
Inlet connectionNW25
Exhaust connectionNW16
Gas ballast1/4" NPTF (two positions)
Water vapor handling per PNEUROP 6602:❑ Position 1: 20.7 g/hr ❑ Position 2: 235 g/hr
Ambient operating temperature5 °C to 40 °C (41 °F to 104 °F)
Storage temperature–20 °C to 60 °C (–4 °F to 140 °F)
Motor rating0.75 HP (550 W)
Operating voltages❑ 1 phase / 100-120 (± 10%) VAC, 50/60Hz ❑ 1 phase/ 200-240 (± 10%) VAC, 50/60 Hz
Run currentSee Table 3 on page 25.
Motor thermal protectionAutomatic thermal protection
Operating speed60 Hz: 1725 RPM, 50 Hz: 1450 RPM
Cooling systemAir-cooled
WeightPump only: 30 kg (66 lbs) Shipping weight: 54 kg (118 lbs)
Leak rate (with exhaust and gas ballast sealed)<1 x 106 sccs helium

Table 1 Specifications (Continued)

CharacteristicValue
Vibration level at inlet, per ISO 10816-19 mms-1 rms
Noise level, per ISO 1120150 dBA ± 2dBA
Fuse typeLittelfuse, Series 215, 10 A, 250 V, 20 mm L x 5.2 mm D, Time Delay / Slow Blow or equivalent, vendor part number 0215010 P.
PollutionDegree 2
InstallationCategory II
Altitude2000 m
Humidity0- 95% (non-condensing)
Indoor use only

Table 2 IDP-15 Power Cord Selection

CountryPower CordPart Number
Europe10 A / 220-230 VAC, 2.5 m656494220
Denmark10 A / 220-230 VAC, 2.5 m656494225
Switzerland10 A / 230 VAC, 2.5 m656494235
UK/Ireland13 A / 230 VAC, 2.5 m656494250
India10 A / 220-250 VAC, 2.5 m656494245
Israel10 A / 230 VAC, 2.5 m656494230
Japan12 A / 100 VAC, 2.3 m656494240
North America15 A / 125 VAC, 2.0 m656458203
10 A / 230 VAC, 2.5 m656494255

RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION

REGIONCONTACT DETAILS
EUROPEFax: 00 39 011 9979 330 Fax Free: 00 800 345 345 00 Toll Free: 00 800 234 234 00 vpt-customercare@agilent.com
NORTH AMERICAFax: 1 781 860 9252 Toll Free: 800 882 7426, Option 3 vpl-ra@agilent.com
PACIFIC RIMplease visit our website for individual office information http://www.agilent.com

Table 3 — IDP-15 Typical Full Load Run Currents (A)

Frequency100 V Nominal120 V Nominal200 V Nominal240 V Nominal100 V120 V200 V230 V240 V
60 Hz8.057.094.013.603.54
50 Hz7.807.933.903.903.94

Table 5 Sélection de câble d'alimentation DP-15

PaysCâble électriqueNuméro de pièce
Europe10 A / 220-230 VAC, 2.5 m656494220
Danemark10 A / 220-230 VAC, 2.5 m656494225
Suisse10 A / 230 VAC, 2.5 m656494235
Royaume-Uni/Irlande13 A / 230 VAC, 2.5 m656494250
Inde10 A / 220-250 VAC, 2.5 m656494245
Israël10 A / 230 VAC, 2.5 m656494230
Japon12 A / 100 VAC, 2.3 m656494240
Amérique du Nord15 A / 125 VAC, 2.0 m656458203
10 A / 230 VAC, 2.5 m656494255

Table 6 — Courants d'exécution de charge pleine typiques (A) IDP-15

Fréquence100 V Nominale120 V Nominale200 V Nominale240 V Nominale100 V120 V200 V230 V240 V
60 Hz8.057.094.013.603.54
50 Hz7.807.933.903.903.94

Table 8 Troubleshooting Chart

ProblemPossible CausesCorrective Actions
Will not startMotor thermal protector openAllow motor to cool. Identify cause of overload.
Blown fuseReplace fuse.
Excessive voltage dropCheck size and length of cable.
Defective motorInspect. Contact Agilent.
Poor ultimate pressureSystem leakLocate and repair leak.
Water in pumpFlush pump with air or dry nitrogen.
Solvent in pumpFlush pump with air or dry nitrogen. Install trap or filter.
Seals worn outReplace tip seals.
Poor conductance to pumpReplumb with shorter and/or larger diameter tubing.
Integrated Isolation Valve (option) solenoid failureContact Agilent.
Hammering noisePump overheatedCheck ventilation to pump. Check ambient temperature.
Debris in pumpCheck intake screen. Flush pump. Disassemble pump and inspect.
Pump runs intermittentlyMotor thermal protector is cycling open and closed.*Allow the motor to cool. Identify the cause of the overload.

* The IDP-15 is equipped with an auto-reset thermal motor protector. This protector automatically shuts down the pump when it detects an overload condition, and automatically restarts the pump when the motor has cooled to within an acceptable temperature range.

Table 9 IDP-15 Service Kits and Advance Exchange Pumps

Part NumberDescriptionContents
X3815-67000Replacement Tip Seal SetReplacement Tip Seal, O-ring, and grease
X3815-69000Exchange IDP-15 StandardIDP-15 Exchange Pumps
X3815-69010Exchange IDP-15 Pumps with VPI ValveIDP-15 VPI Valve Exchange Pumps

Request for Return Form (Health and Safety Certification)

FieldValue
1) CUSTOMER INFORMATION — Company Name
Contact Name
Tel
Email
Fax
Customer Ship To
Customer Bill To
Europe only: VAT reg. Number
USA/Canada only: Taxable
USA/Canada only: Non-taxable
2) PRODUCT IDENTIFICATION — Product Description
Agilent P/N
Agilent S/N
Original Purchasing Reference
3) TYPE OF RETURN — 3A. Non-Billable
3A. Billable
3A. New PO # (hard copy must be submitted with this form)
3B. Exchange
3B. Repair
3B. Upgrade
3B. Consignment/Demo
3B. Calibration
3B. Evaluation
3B. Return for Credit
4) HEALTH and SAFETY CERTIFICATION — HAS NOT pumped or been exposed to any toxic or hazardous materials
HAS pumped or been exposed to the following toxic or hazardous materials
Toxic
Corrosive
Reactive
Flammable
Explosive
Biological
Radioactive
List all toxic/hazardous materials. Include product name, chemical name, and chemical symbol or formula
Print Name
Authorized Signature
Date
5) FAILURE INFORMATION — Failure Mode (REQUIRED FIELD. See next page for suggestions of failure terms)
Detailed Description of Malfunction: (Please provide the error message)
Application (system and model)

Форма содержит чекбоксы и поля для заполнения; пустые ячейки оставлены пустыми строками.

Engineering Facts

14 facts · every claim linked to a page
Selection

The pump is only suitable for air or inert gases and must not be used on corrosive, explosive, or particulate-forming gases.

Determines whether the pump can be applied to a given process gas at all.

Source
“This pump is suitable for pumping air or inert gases. The pump is not intended to pump corrosive, explosive, or particulate-forming gases.”
Installation

Shipping locks must be disengaged before operation or the pump will not float on its isolation mounts.

Prevents damage and ensures vibration isolation is active at startup.

Source
“Before operating the IDP-15 disengage the shipping locks.”
Installation

Shipping locks must be re-engaged for transport, long moves, or long-term storage.

Protects the scroll set and isolation mounts during handling and storage.

Source
“Engage the shipping locks (Figure 3) when shipping, moving a long distance or during long-term storage.”
Installation

The voltage selector must be set to match the actual mains supply before powering up.

Wrong setting risks motor damage and invalidates the electrical setup.

Source
“Slide the voltage selector switch to 115 for 100-120 V mains, or to 230 for 200-240 V mains.”
Installation

The pump must be fed from a 15 A circuit at 100/120 V to avoid nuisance breaker trips on start.

Sizing the branch circuit correctly avoids spurious trips during motor start.

Source
“For 100/120 V service, connect the IDP-15 to a 15 A circuit.”
Installation

A minimum 2-inch clearance at front and rear is required to keep fan ducts unobstructed.

Prevents overheating and shortened pump life in enclosures or racks.

Source
“Leave a minimum gap of 2" between front and back of the pump.”
Installation

Exhaust must not be restricted by long small-diameter tubing to avoid overheating.

Guides exhaust plumbing diameter and length during system integration.

Source
“To avoid overheating the pump, do not restrict the exhaust flow with long lengths of small diameter tubing.”
Installation

A bellows between pump and chamber provides both vibration isolation and strain relief.

Protects the chamber and piping from pump vibration and mechanical stress.

Source
“Insert a bellows between the pump and vacuum chamber to provide both vibration isolation and strain relief.”
Operation

The pump needs about an hour of running with the inlet sealed to reach ultimate vacuum.

Sets realistic expectations for commissioning and acceptance testing.

Source
“In order to reach maximum vacuum, the pump must be left running for about an hour with the inlet sealed.”
Operation

Gas ballast port 2 handles high water vapor loads while port 1 suits moderate loads.

Selecting the correct ballast port prevents excessive tip seal wear.

Source
“Gas ballast port 2 is used for high water vapor loads”
Operation

For air-sensitive applications dry nitrogen at about 5 l/min can be fed to the gas ballast inlet.

Allows inert ballast purge to avoid air ingress in sensitive processes.

Source
“dry nitrogen at a flow rate of approximately 5 l/min can be provided to the gas ballast inlet”
Maintenance

The pump has no oil for cleaning, so periodic short runs with the inlet open flush out debris.

Defines a simple preventive maintenance routine for contaminated processes.

Source
“Run the pump periodically with the inlet open to atmosphere for a minute or two to flush it out.”
Maintenance

Newly installed tip seals may need several hours of run time to seat and meet specifications.

Prevents premature rejection of a rebuilt pump during performance checks.

Source
“Newly installed tip seals may require several hours of run time to seat properly”
Safety

The gas ballast must be sealed whenever pumping gas not intended to be vented to atmosphere.

Avoids uncontrolled release of hazardous process gas to the room.

Source
“The gas ballast must be sealed whenever pumping any gas not intended to be vented to the atmosphere”