Induztr
Resumen técnico Visor de documentos Modelos relacionados

IDP-4 Dry Scroll Pump User Manual

Contenido

  1. Overview pp. 3-4
    Table of contents listing manual sections
  2. Spec pp. 5-6
    Environmental, conformity and utility requirements
  3. Procedure pp. 7-8
    Steps to prepare pump for installation
  4. Overview pp. 9-10
    Intended use and operating limits
  5. Spec pp. 11-12
    Technical data and specifications
  6. pp. 13-38
    Full manual translated into Italian
  7. pp. 39-62
    Full manual translated into German
  8. pp. 63-88
    Full manual translated into French
  9. pp. 89-112
    Full manual translated into Spanish
  10. pp. 113-138
    Full manual translated into Traditional Chinese
  11. pp. 139-162
    Full manual translated into Japanese
  12. pp. 163-186
    Full manual in English
  13. Spec pp. 187-190
    Detailed technical data and specifications
  14. Spec pp. 191-192
    Dimensional outline drawing of pump
  15. Spec pp. 193-194
    Pumping speed and performance curves
  16. Procedure pp. 195-196
    Electrical wiring and connection instructions
  17. Procedure pp. 197-198
    Inlet and outlet mechanical connections
  18. Procedure pp. 199-200
    Gas ballast operation and adjustment
  19. Ordering pp. 201-202
    List of accessories and spare parts
  20. Procedure pp. 203-206
    Cleaning and preventive maintenance procedures
  21. Procedure pp. 207-208
    Procedure for replacing tip seals
  22. Services pp. 217-218
    Services requiring factory return
  23. Procedure pp. 219-220
    Troubleshooting guide and remedies
  24. Spec pp. 221-222
    Serial connector pin allocation
  25. Spec pp. 223-226
    Communication window description and protocol
  26. pp. 227-230
    Vacuum Products Division contact details
  27. p. 189 table, 17 rows
  28. p. 187 table, 28 rows
  29. p. 190 table, 9 rows
  30. p. 195 table, 9 rows
  31. p. 216 table, 11 rows
  32. p. 224 table, 19 rows
  33. p. 203 table, 11 rows
  34. p. 201 table, 5 rows
  35. p. 207 table, 4 rows

Table 1 IDP-4 Technical Specifications

ModelIDP-4 Dry Scroll Pump
Interface dimensionsH=181 mm x W=139 mm x L= 288 mm (7.1X5.5X11.3”)standard pump H=237 mm x W=139 mm x L= 288 mm (VPI Valve) H=237 mm x W=139 mm x L= 338 mm (with all accessories)
Peak pumping speed (ISO 21360-1: 2020) (ISO 21360-2: 2020)65 L/min, (4 m3/h, 2.35 cfm) at full speed setting Includes average peak pumping speed for both N2 and He without inlet valve configuration
MediaNo corrosive, explosive or particulate forming gases
Ultimate Pressure* (ISO 21360-1: 2020) (ISO 21360-2: 2020)1.8 +/- 0.2 x 10-1 Torr (2.4 +/- 0.27 x 10-1 mbar) at 3300 rpm (default setting)
Leak Rate<1 x 10-6 sccs helium
Maximum recommended continuous inlet pressure1.0 atmosphere (0 psig)
Maximum inlet pressure at nominal voltage1.0 atmosphere (0 psig)
Maximum outlet pressure***0.45 bar (6.5 psig)
Inlet connectionNW16
Exhaust connectionNW16
Gas ballastFemale 1/8" National Pipe Thread (adapter provided)
Ambient operating temperature5 °C to 40 °C (41 °F to 104 °F)
Ambient Relative Humidity5-95%; non-condensing
Storage temperature–20 °C to 60 °C (–4 °F to 140 °F)
Motor rating: EN 60034-1-2:2007 EN 60034-30 (GB 18613-2020)0.16 HP (0.12Kw)
Operating voltages** Single phase100-127VAC 50/60Hz 200-240 50/60Hz
Motor protectionThermal protection/over current protection/fuse

* Ultimate Pressure; ** Operating voltages; *** Maximum outlet pressure

Technical Information

РазделСтраница
Description of the IDP-4 Dry Scroll Pump188
Technical Specifications189
IDP-4 Dry Scroll Pump Outline Drawing191
Performance Curves193
Electrical Connections195
Grounding Instructions195
Mechanical Connections197
Gas Ballast199
Accessories and Spare Parts201
Optional Accessories201
Optional Inlet Valve Kit202
Installation204
Cleaning and Preventive Maintenance205
External cleaning205
Internal cleaning as part of a Preventive Maintenance plan206
Verifying Performance206
Tip Seal Replacement207
Tip Seal Replacement Instructions208
User Maintenance and Repair216
Factory-Only Service217
Service Options218
Troubleshooting219
Serial Connector & Pin allocation221
Pump Serial Connector221
Serial communication settings222
Window Description and Protocol223
Window Description223
Windows Protocol226

Table 1 IDP-4 Technical Specifications (continued)

ModelIDP-4 Dry Scroll Pump
Operating speedVariable Speed (2200-3300rpm), default setting 3300rpm +/-10%
Cooling systemAir-cooled
WeightPump only: 7.7kg (17 lbs) Shipping weight: Approximately 9.8 kg (22 lbs) with accessories
Sound Pressure Levels (EN ISO 2151:2008) (ISO 4871:1996) (ISO 11201:2010)50 to 55 +/- 2.5 dba (min to max RPM setting) with default configuration (gas ballast closed)
Fuse type5A slow blown 250V 5X20 mm
InstallationIndoor use, Equipment Class 1, Overvoltage Category II, Pollution Degree 2
Altitude2000 m (6562 ft)
Emissions and ImmunityElectromagnetic Compatibility 2014/30/EU:2014 EN 61326-1: (2013) as a Class A device EN 61326-1: (2013) as a Class A device EN 61000-3-2:2019 Class A EN 61000-3-3:2014/A2:2022 Class A EN 61000-6-2:2020
SafetyMachinery Directive 2006/42/EC: 2006 EN 1012-2:1996+A1: 2010 EN ISO 12100:2012 Low Voltage Directive 2006/95/EC: 2007 EC 61010-1 Ed. 3.1 b:2018 CSA certified for US and Canada CE, ROHS

* If the pump is new or has been exposed to atmosphere for an extended period of time, it is possible that water can be present which can affect the pump’s ability to achieve ultimate pressure. To remove water from the pump and system, activate the gas ballast using the procedure in the gas ballast paragraph on page 199. Ultimate pressure will vary for different configuration and RPM setting. ** Mains supply voltage fluctuations shall not exceed ±10% of the nominal voltage. For operation in Japan, the pump will function at 100VAC, but may offer reduced peak pumping speed at maximum load conditions. For operation in Japan at high inlet pressures and high pumping loads, use of 200VAC is recommended. *** For customers with outlet pressure might experience different sound quality and slower pumping speed.

Table 2 Standard IDP-4 Power Cord Options

CountryPower Cord SpecificationPart Number
Europe10 A / 220-230 VAC, 2.5 m656494220
Denmark10 A / 220-230 VAC, 2.5 m656494225
Switzerland10 A / 230 VAC, 2.5 m656494235
UK/Ireland13 A / 230 VAC, 2.5 m656494250
India10 A / 220-250 VAC, 2.5 m656494245
Israel10 A / 230 VAC, 2.5 m656494230
Japan12 A / 100 VAC, 2.3 m656494240
North America10 A / 230 VAC, with C13 2.5 m (NEMA 6-15)656494255
North America13 A / 125 VAC, 2.0 m656458203

Table 7 Maintenance Intervals and Part Numbers

Service OperationIntervalParts NeededPart NumberDescription
Regular Maintenance
Gas Ballast ActivationSee Table 5 for suggested intervals based on applicationNone
Vent pump to atmospheric pressureNone
Minor Maintenance: Tip Seal & Main O-ring replacementIDP3TSIDP-3, 4 Tip Seal Kit
Replace Exhaust filter elementYearly, if exhaust silencer is installed.REPLSLRFILTER2FILTER ELEMENT, IDP3,4 EXHAUST
If-Required Component Replacement
Front Cowling and cooling fanIf damaged or fan not functioningX3802-67300IDP-4 Fan & Cowling Kit
FeetIf broken or wornIDP3VIBISOKITIDP-3, 4 Vibration Isol. Feet Kit
Plastic exhaust ManifoldIf cracked, broken, or leaking.X3803-67005IDP-3, 4 Std Exhaust mfld Kit
VPI ValveIf leaking or damagedVPI16IDP24DCSub-Assy, VPI Valve-24VDC, IDP3, 4
Standard Exhaust FlangeIf damagedX3802-67120IDP-4 Std Exhaust Flange Kit

Table 11 Window meanings (continued)

Win #ReadWriteTypeNameMeaningDefault (Factory Set)Unit
225AlwaysAlwaysLInlet valve controlControl inlet valve open or close command This function is disabled during the 10-second startup delay that occurs when power is supplied to the pump or the power switch is turned ON.--
227AlwaysAlwaysLGas ballast valve controlControl gas ballast open or close command--
232AlwaysOffNRotational FrequencyRotational Reading Hz-0.1Hz
233AlwaysOffNRotational FrequencyRotational Reading RPM-RPM
234AlwaysOffNBus VoltageIt reads the Bus Voltage-Volt
236AlwaysOffNTorqueEstimate Mechanical Torque-N-mm
300AlwaysOffNCycle timeIt indicates the time of the last complete cycle power on to power off-Minutes
301AlwaysOffNCycle numberIt indicates the number of cycles start/stop during all the pump life.--
302AlwaysOffNPump lifeIt indicates the total hours of pump life. The motor is running.-Hours
303AlwaysOffNTotal Error CountsTotal errors counts for life.--
319AlwaysAlwaysAController Part NumberIt is the Part Number of the controller--
320AlwaysAlwaysAPump Part NumberIt is the Part Number of the pump--
321AlwaysAlwaysAPump Special ModelIt indicates the Part Number of special pumps--
322AlwaysAlwaysAPump Serial NumberIt indicates the Serial Number of the pump--
323AlwaysAlwaysAController Serial NumberIt indicates the Serial Number of the controller--
324AlwaysAlwaysAPump Modification LevelIt indicates the upgrading level of the pump--
325AlwaysAlwaysAController Modification LevelIt indicates the upgrading level of the controller--
358AlwaysAlwaysNPump life since last serviceIt indicates the days after the last resetting-Days
399AlwaysOffNController lifeIt indicates the total hours of Controller life. The total time the controller is powered on.-Hours

Table 4

ParameterValue
Leak Rate:
Body< 1X10-9 atm cc/sec He
Seal<1X10-9 atm cc/sec He
Closing time (after solenoid valve power loss)25 ms
Closing burst0.01 torr-liter
Materials:
BodyAluminum (6061-T6)
SealViton
Temperature Range:
Valve Body0~100 °C
Solenoid0~50 °C

Table 3 Accessories

Part NumberDescription
IDP3GBKITGas Ballast kit
VPI16IDP24DCVPI Valve-24VDC, IDP-4
EXSLRIDP3Exhaust Silencer Kit
IDP3GBKITBreather Vent and Adapter
X3802-68001Serial communication cable kit

Table 5 Suggested maintenance based on duty level

DescriptionLight-DutyMedium-DutyHeavy-Duty
Intermittent use Pump is shut down between experiments.Continuous low-pressure use. Ex. turbo backing Low gas loadContinuous high inlet pressure use. High gas load
Vent to atmosphereUpon start up and preparation for the experiment, vent the inlet to atmosphere for 5 seconds once the pump has reached full speed. Ensure the exhaust is not restricted. 1x per month if the application permits.Properly isolate any sensitive equipment prior to venting. Run the pump at full speed with the inlet fully open for 5 seconds. Ensure the exhaust is not restricted. 1x per yearProperly isolate any sensitive equipment prior to venting. Run the pump at full speed with the inlet fully open for 5 seconds. Ensure the exhaust is not restricted.
Gas Ballast UseUpon start up and preparation for the experiment, run the pump with the inlet capped and the gas ballast open for 4 hours to dry out any water accumulated during a preceding idle period.Continuous gas ballast use is recommended if the application permits.Periodic use of the gas ballast is likely required. Continuous use of the gas ballast is recommended if pumping a steady load of water vapor.
Tip Seal ReplacementPerform a minor maintenance on the pump every 10,000 hours, or when the pump’s ultimate pressure cannot be reached with overnight gas ballast use.Perform minor maintenance on the pump every 18 - 24 months. Ensure that performance degradation is not due to water vapor prior to completing maintenance.Perform minor maintenance once per year.

Engineering Facts

11 facts · every claim linked to a page
Selection

The pump is only for indoor use in clean, dry environments and is not certified for outdoor operation.

Determines whether the pump can be specified for a given site and enclosure class.

Source
“The IDP-4 Dry Scroll pump is designed and certified for operation indoors (installation category II) in relatively clean, dry environments (pollution degree 2).”
Selection

The pump is not intrinsically safe and must never be used with corrosive, pyrophoric, particle-forming or flammable gas mixtures.

Rules out the pump for aggressive or hazardous process gases at the selection stage.

Source
“The IDP-4 pump is NOT intrinsically safe and should never be used with corrosive, pyrophoric, particulate-forming gases or gas mixtures in their flammable and/or combustible range.”
Installation

The power cord must be IEC-320 style rated at least 10 A.

Defines the minimum supply cord rating needed for a compliant installation.

Source
“Connect the pump to the power supply using an IEC-320 style power cord of at least 10 A capacity.”
Installation

Operating ambient must stay within +5 to +40 °C and 5–95% non-condensing humidity.

Sets the environmental envelope the installation must satisfy.

Source
“Temperature: +5 °C to +40 °C (41 °F to 104°F)”
Installation

The pump needs at least 20 cm clearance on all sides for ventilation, maintenance and switch access.

Dictates the physical layout and free space around the pump.

Source
“This pump requires a clear space of at least 20 cm (8 in.) on all sides for ventilation, maintenance access and easy access to the power switch.”
Installation

The power switch and cord connection must remain easily accessible for quick disconnection.

Affects placement so the pump can be de-energized quickly in an emergency.

Source
“it is recommended that the pump be positioned in a way that the power switch and power cord connection remain easily accessible to the user in order to disconnect power pump quickly if needed.”
Operation

Poor initial vacuum is usually adsorbed water vapor; gas ballast for ~30 min to ~24 h restores full performance.

Explains expected pump-down behavior and the remedy at commissioning.

Source
“it can typically take between ~30 minutes to ~24 hours of gas ballast activation to achieve the full vacuum performance.”
Operation

Gas ballast must not be used with hazardous materials or gases that must not mix with air.

Prevents unsafe venting of process gases to atmosphere.

Source
“Do not use gas ballast if using potentially hazardous materials or those that should not be mixed with air.”
Operation

Blocking cooling vents can overheat the pump; surface above 65 °C is potentially damaging.

Defines the overheating threshold and the need for unobstructed airflow.

Source
“A pump surface temperature in excess of 65 °C (150 °F) is potentially damaging.”
Maintenance

Wait for the pump to fully stop and cool to room temperature before servicing or disconnecting.

Protects personnel from sharp edges and hot surfaces during service.

Source
“Before servicing the pump or before any mounting or disconnecting of the pump from the system, wait for the pump to come to a complete stop and cool down to room temperature.”
Maintenance

Equipment heavier than 20 kg must be moved with a suitable lifting device.

Sets the manual-handling limit for transport and installation.

Source
“Any equipment heavier than 20 kg should be transported using a suitable lifting device.”

Modelos relacionados